Checks/tests
Operating manual AVERO Motion
V2.6 E
Page 27 of 48
10 Checks/tests
In order to ensure a safe use of our AVERO Motion bath tub and the protection of the
users and the patients, BEKA Hospitec prescribes an annual safety check.
The execution of the safety checks and maintenance must be documented and proven
on request. Please use your inventory register to this effect.
We recommend a simultaneous maintenance of the device in order to conserve its full
value.
The checks may only be carried out by appropriately trained and qualified personnel.
Failure to do so may result in injury or a non-safe product.
NOTE
In accordance with the UVV (accident prevention) regulations of
the German employer’s liability insurance association on
stationary equipment which is used in special locations or
installations, an annual check to the DGUV (German Statutory
Accident Insurance Association) Prescription 3 (BGV A3) must be
carried out on the AVERO Motion bath tub .
This check is only prescribed for Germany. In other countries, other requirements
might be applicable.
WARNING
Do not conduct any cleaning, maintenance or test activities when
the bath tub is in use. This could cause danger to the user and
the patient.
measures to be conduct
Before each
application /
daily
weekly monthly
yearly
Check the level of the disinfectant. (If
available)
X
Clean and disinfect the tub.
X
Check the tub for function.
X
Conduct a visual inspection of all
components, power cables, hoses and
connections.
X
Check the water inlet and clean it if
necessary.
X
We recommend regular maintenance, a
safety check and a check to DGUV
prescription 3.
X
We recommend to replace
biennial
the batteries of optional emergency drain function.
Summary of Contents for AVERO Motion
Page 34: ...Appendix Operating manual AVERO Motion V2 6 E Page 34 of 48 13 3 2 Connection at the rear...
Page 35: ...Appendix Operating manual AVERO Motion V2 6 E Page 35 of 48 13 3 3 Connection at the floor...
Page 36: ...Appendix Operating manual AVERO Motion V2 6 E Page 36 of 48 13 4 Declaration of Conformity...
Page 47: ...Appendix Operating manual AVERO Motion V2 6 E Page 47 of 48 Your notes...
Page 48: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...