Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E
Seite
19
von
58
Switch on the water supply via hand shower
or
via the water inlet
Figure 15:
Adjust the water to the
desired maximum
discharge
temperature. Measure
the temperature with
the thermometer
Figure 16:
The picture shows the
desired maximum
temperature, which
must agree with the
manually measured
temperature.
Figure 17:
Slide the knob to the
maximum position.
Figure 18:
Switch off the solenoid
valve.
Figure 19:
Install the rotary knob
with the cross bolt
Figure 20:
Reinsert the cover.
Figure 21:
Switch on a solenoid
valve (hand shower or
water spout) to check
the setting again.
Figure 22:
Check the outlet
temperature with the
external thermometer!
The process is now
complete.
41,5