15.8 Electromagnetic compatibility
Electrical medical equipment needs to be installed with special precautions taken regarding
EMC and must be installed and put into operation according to EMC information contained in
the accompanying documents.
No special measures must be taken for devices and systems of BEKA Hospitec Ltd.
Portable and mobile RF communications equipment can affect electrical medical equipment.
Guidelines and manufacturer declaration - electromagnetic immunity
The AVERO Premium Plus is intended for use in the ELECTROMAGNETIC ENVIRONMENT
specified below. The customer or user of the AVERO Premium Plus should ensure that it is used
in such an environment.
Release measurements
Agreement
Electromagnetic environment - guidelines
High-frequency (RF) emissions
according to
CISPR 11
Group 1
The AVERO Premium Plus uses RF energy for its
internal functions only. Therefore, the RF radiation of
the device is very low and it is unlikely that nearby
electrical devices will be disturbed.
High-frequency (RF) emissions
according to
CISPR 11
Class B
The AVERO Premium Plus is intended for use in all
facilities including residential areas and the like that are
directly connected to a public power supply that also
supplies buildings used for domestic purposes.
Harmonics according to
IEC 61000-3-2
Class A
Voltage fluctuations /
flickers according to
IEC 61000-3-3
complies