Appendix
AVERO Premium Plus User Manual
V2.4 E
- Page Page
51
of
58
Guidelines and manufacturer declaration - electromagnetic immunity
The AVERO Premium Plus is intended for use in the ELECTROMAGNETIC ENVIRONMENT
specified below. The customer or user of AVERO Premium Plus should ensure that it is used in
such an environment.
Interference immunity
tests
IEC 60601- test level
Consistency
level
Electromagnetic
Environment
Guidelines
Static discharge
Electricity (ESD) according
to
IEC 61000-4-2
± 6kV Contact discharge
± 8kV Air discharge
± 6kV Contact discharge
± 8kV Air discharge
The floor should be
wood, concrete or
ceramic tile. If floors are
covered with synthetic
material, the relevant
humidity should be at
least 30%.
Fast transient
interference pulses / Burst
IEC 61000-4-4
± 2 kV for
Power supply lines
± 1 kV for
Input /
output lines
± 2 kV for
Power supply lines
Not applicable for input /
output lines
The quality of the
power supply should be
that of a typical
commercial or hospital
environment.
Overload
IEC 61000-4-5
± / to 1 kV line opposite
line
± / to 2 kV line opposite
ground
± / to 1 kV line opposite
line
± / to 2 kV line opposite
ground
The quality of the
power supply should be
that of a typical
commercial or hospital
environment.
Voltage dips,
short interruptions
and
voltage fluctuations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
)
for 0.5 period
40 % U
T
(>60 % dip in U
T
)
for 5 periods
70 % U
T
(>30 % dip in U
T
)
for 25 periods
<5 % U
T
(>95 % dip in U
T
)
for 5 s
<5 % U
T
(>95 % dip in U
T
)
for 0.5 period
40 % U
T
(60 % dip in U
T
) for
5 periods
70 % U
T
(30 % dip in U
T
) for
25 periods
<5 % U
T
(>95 % dip in U
T
)
for 5 s
The quality of the
power supply should be
that of a typical
commercial or hospital
environment.
Power Frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Magnetic fields at mains
frequency should have
the typical values that are
found in a business or
hospital
environment.
ATTENTION
U
T
is the ac mains voltage prior to application of the test level.