Unpacking and assembly
User Manual AVERO Premium Plus V2.4 E
Seite
17
von
58
4.6
Thermal rinse and subsequent adjustment of the outlet temperature
Note: Follow the steps carefully. If not, the disinfecting device may be blocked.
You can only remove this blocking by disconnecting from the mains voltage!
Figure 1:
Tools you need:
1 x screwdriver flat,
0,6 x 3,5 mm,
1 Screwdriver Cross-
hatch,
1 x thermometer
Figure 2:
Turn the knob to the
left until it stops (Snap
button not pressed)
Figure 3:
Use the screwdriver to
remove the cap
Figure 4:
Unscrew the cross
screw from the rotary
actuator and pull it
forwards.
Figure 5:
Open the solenoid
valve of the hand
shower
Figure 6:
Turn the knurled knob
to the left to increase
the temperature
Figure 7:
If the temperature
exceeds 42 ° C, the
control switches off for
safety reasons. The
display shows "hot"
alternating with the
temperature.
Figure 8:
Press and hold
button 5, than press
the buttons 8 and 10
together
This bridges the
protection circuit and
allows you to set a
higher temperature.