Operating
User Manual AVERO Premium Plus V2.4 E
Seite
33
von
58
10.9 Emergency drain function during power failure (option)
If have you have integrated the
“emergency drain”
option on your tub, the tub is supplied with
an internal battery in the case of a power failure. The batteries only supply power to the control
panel, the hand control panel and motor for lifting and lowering the tub. All other functions (i.e.
air spa, light therapy etc.) will shut down in the case of a power failure.
Note:
The emergency drain function is deactivated to protect the
internal battery from exhaustive discharge upon delivery of
the AVERO Premium Plus.
To use the
“emergency” drain option, the switch must be
turned to “ON”.
The switch for activating the emergency drain function is located under the removable
inspection cover for refilling the disinfectant.
Note:
The battery capacity is designed so that you can move the
tub once to the lower position.
This feature allows you to help the patient out of the tub
safely even during a power failure.
If the battery capacity is exhausted none of the processes
will function, even if there is no power failure.
In a normal state the internal batteries will be at a conservation verify again held by an
intelligent charging circuit.
In case of a power failure “Accu” will appear on the control panel and the
temperature indicators will switch off. Add-on modules such as air spa, light
therapy etc. will automatically switch off.
Move the tub to the lowest position and help the patient out of the tub, or use
a lift system to lift the patient out of the tub.