Appendix
Operating manual SENTA PUR L
V2.5 E
Page 58 of 64
Guidance and manufacturer’s declaration - electromagnetic immunity (Table 202)
The SENTA PUR L has been designed for use in the hereafter listed ELECTROMAGNETIC
ENVIRONMENTS. The customer or the user of the SENTA PUR L must ensure that the
appliance is used in such environment.
Immunity
testing
IEC 60601- Test level
Compliance level
Electromagnetic
environment guidance
Discharging of
static electricity
(ESD) to IEC
61000-4-2
± 6kV contact discharge
± 8kV air discharge
± 6kV contact
discharge
± 8kV air discharge
The floor must be in wood,
concrete or ceramic tiles. In
case of floors in synthetic
material, the relevant air
humidity must be at least
30%.
Rapid transient
interference
pulses/burst
IEC 61000-4-4
± 2 kV for power supply
cables
± 1 kV for input/output
cables
± 2 kV for power
supply cables
not applicable to input/
output cables
The quality of the supply
voltage should match that of
a typical business or hospital
environment.
Overvoltage
IEC 61000-4-5
±/1 kV cable against
cable
±/2 kV cable against
ground connection
±/1 kV cable against
cable
±/2 kV cable against
ground connection
The quality of the supply
voltage should match that of
a typical business or hospital
environment.
Voltage drops,
short interruptions
and voltage
fluctuations in the
power supply
input cables
IEC 61000-4-11
<5 % U
T
(>95 % drop of U
T
)
for 0.5 period
<40 % U
T
(>60 % drop of U
T
)
for 5 periods
<70 % U
T
(>30 % drop of U
T
)
for 25 periods
<5 % U
T
(>95 % drop of U
T
)
for 5 s
<5 % U
T
(>95 % drop of U
T
)
for 0.5 period
<40 % U
T
(>60 % drop of U
T
) for
5 periods
<70 % U
T
(>30 % drop of U
T
) for
25 periods
<5 % U
T
(>95 % drop of U
T
)
for 5s
The quality of the supply
voltage should match that of
a typical business or hospital
environment.
Current frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical business
or hospital environment.
CAUTION U
T
is the mains AC voltage before the application of the test level.
Summary of Contents for SENTA PUR L
Page 54: ...Appendix Operating manual SENTA PUR L V2 5 E Page 54 of 64 14 2 Dimensions of the SENTA PUR L...
Page 55: ...Appendix Operating manual SENTA PUR L V2 5 E Page 55 of 64 14 3 Declaration of conformity...
Page 63: ...Appendix Operating manual SENTA PUR L V2 5 E Page 63 of 64 Your notes...
Page 64: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...