Operation
Operating manual SENTA PUR L
V2.5 E
Page 40 of 64
In
LIVE WT
mode (weighing of persons), proceed as follows:
If the device is within the zero range, the weight is displayed normally,
i.e. the Live-WT function is deactivated.
Press the
<TARE>
or
<ZERO>
button to set the emptied scale to
zero.
Put the person on the scale.
As soon as the weight is outside of the zero range, an average is
calculated over a longer period of time. After that, the device displays
Hold
, the weight display is “frozen” and the device emits a beep.
Press the
<LIVE WT>
button once more to trigger a new average
calculation.
As soon as the gross weight is within the zero range, i.e. the scale is
emptied, the process can be repeated.
8.12.13
Error message in operation
Error
Description
Measure
(U - - - -)
Scale underload
Increase weight or switch
scale off and back on.
(O - - -- -)
Scale overload
Reduce weight.
(TARE)
(ERROR)
Taring trial outside of authorised range.
(Calibrated scales must only be tarred
e.g. in case of a positive gross weight).
Change weight.
(ZERO)
(ERROR)
Zeroing trial outside of authorised range.
Change weight or switch
scale off and back on.
(STABLE)
(ERROR)
A
<ZERO>
,
<TARE>
or
<PRINT>
command cannot be executed in the
absence of a standstill.
Repeat the procedure when
the scale is stable.
Summary of Contents for SENTA PUR L
Page 54: ...Appendix Operating manual SENTA PUR L V2 5 E Page 54 of 64 14 2 Dimensions of the SENTA PUR L...
Page 55: ...Appendix Operating manual SENTA PUR L V2 5 E Page 55 of 64 14 3 Declaration of conformity...
Page 63: ...Appendix Operating manual SENTA PUR L V2 5 E Page 63 of 64 Your notes...
Page 64: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...