Operation
Operating manual SENTA PUR L
V2.5 E
Page 42 of 64
8.12.15
Diagnostic error messages
Error
Description
Remedy
E0001
Voltage supply too low
Check
E0002
Voltage supply too high
Check
E0010
Temperature too high/low
(-10 to +50
C ambient).
Check local conditions
E0020
Incorrect scale division
(target between 100 and 30000 parts).
Change value
E0100
Settings are lost
Repeat settings
E0200
Scale adjustment is lost
Re-calibration
E0300
All settings and scale adjustment are lost.
Completely reconfigure
settings and calibrate
E0400
Factory settings are lost (FATAL).
Service
E0800
EEPROM defect (FATAL).
Service
E2000
Wrong power supply to weighing cells.
Check weighing cell cable
E4000
Battery-buffered memory lost.
Repeat settings
E8000
FLASH memory defect (FATAL).
Service
The E
– error messages are additive. Example:
•
E0012 (E0010+E0002) indicates that both the voltage supply is too high and the
temperature too high.
•
E0C00 (E0800+E0400) indicates that the factory settings are lost and the EEPROM
is defect.