Operation
Operating manual SENTA PUR L
V2.5 E
Page 43 of 64
8.12.16
Charging the battery of the calibrated patient scale:
The patient scale is supplied with voltage through an 8-cell nickel/metal hydride battery
pack.
Loading process:
1. Plug the charger in the socket.
2. Plug the DC connector of the charger in the charging socket, on the right-hand
side underneath the scale cover of the SENTA lift.
3. The charging indicator will become red, the battery is being charged. The
charging time for a completely discharged battery is 16 hours. (The capacity of
the battery is 2700 mAh. This is sufficient for approx. 24 hours of weighing.)
4. The charging process is not automatically interrupted; therefore, you should
terminate the charging process after the specified time.