background image

The indicator’s digital display at which the bargraph
starts is defined by the ‘bArLo’ sub-function which is
selected by pressing the 

 button followed by the  

P

button which will reveal the current indicator display at
which the bargraph starts.   The flashing digit can be
adjusted by pressing the 

 or 

 buttons, when set as

required  pressing  

P

 

will  transfer  control  to  the  next

digit.  When all the digits have been adjusted, press 

E

to  return  to  the  'bArLo'  prompt  from  which  ‘bArHi’
which defines the finishing point of the bargraph can
be  selected  by  pressing  the  

 button.     ‘bArHi’ is

adjusted in the same way as ‘bArLo’.   When set as
required, pressing 

E

 twice will return the display to the

‘bAr’ prompt in the configuration menu.

Note:

 ‘bArLo’ must be set lower than ‘bArHi’, incorrect

setting is indicated by the bargraph scale flashing with
a single bargraph segment activated.

6.8  Function of the 

P

 push button:   ‘C - - P’

When the indicator is in the display mode, operating
the  

P

  push-button  will  display  the  input  current  in

milliamps, or the displayed value as a percentage of
the difference between the displayed values at 4mA
and 20mA inputs.

To check or change the function of the 

P

 push-button

select 'C - -P' from the configuration menu and press

P

 to reveal the current setting.  Pressing the 

 or 

button  will  toggle  the  setting  between  '4-20'  the
current display in milliamps and 'PC' the percentage
display.   When set as required press  

E

 to return to

the ‘C - - P’ prompt in the configuration menu.

6.9  Tare function:   ‘tArE’

The  tare  function  is  primarily  intended  for  use  with
weighing system.  When the indicator is in the display
mode and the tare function is activated, pressing the

E

 button  for more than  three seconds will  zero the

indicator’s  digital  display  and  activate  the  tare
annunciator.    On  models  with  a  bargraph,  the
bargraph  remains  linked  to  the digital  display when
the tare function is activated.   Subsequent operation
of the 

E

 push button for less than 3 seconds will return

the indicator to the gross display and deactivate the
tare annunciator.

To  check  or  change  the  tare  function  select  'tARE'
from the configuration menu and press 

P

 to reveal the

current  setting.    Pressing  the  

 or  

 button  will

toggle the setting between 'on' and 'oFF'.   When set
as required press  

E

 to return to the ‘tARE’ prompt in

the configuration menu.

6.10  Security code:  ‘CodE’

Access to the instrument configuration menu may be
protected by a four digit security code which must be
entered  to  gain  access.      New  instruments  are
configured with the default security code 0000 which

allows  unrestricted  access  to  all  configuration
functions.

To enter a new security code select 'CodE' from
the  configuration  menu  and  press

 

P

 which  will

cause the indicator to display the existing security
code with one digit flashing.  The flashing digit can
be  adjusted  using  the  

 and

 

 push-buttons,

when set as required operating the  

P

 button will

transfer control  to  the  next digit.   When all  the
digits have been adjusted press

 

E

 to return to the

‘CodE’ prompt. The revised security code will be
activated  when  the  indicator  is  returned  to  the
display  mode.    Please  contact  BEKA associates
sales department if the security code is lost.

6.11  Reset to factory defaults:  ‘rSEt’

This  function  enables  the  indicator  and  the
lineariser  to  be  quickly  returned  to  the  factory
default configurations shown in sections 6 and 7.

To reset the indicator or lineariser select ‘rSEt’ from
the configuration menu and press  

P

, the indicator

will  display  one  of  the  reset  options  ‘ConF’  or
‘LtAb’.

‘ConF’

Resets the indicator to defaults

‘LtAb’

Resets the lineariser to defaults

Using the 

 or 

 push button select the required

sub-function and press  

P.

  To  prevent accidental

resetting  the  request  must  be  confirmed  by
entering ‘5urE’.   Using the 

 button set the first

flashing digit to ‘5’ and press 

P

 to transfer control to

the second digit which should be set to ‘u’.   When
‘5urE’ has been entered pressing the 

E

 button will

reset the selected configuration menus and return
the  display  to  the  ‘rSEt’  function  in  the
configuration menu.

6.12  Under and over-range

If the  numerical  display range of the  indicator is
exceeded,  all  the  decimal  points  will  flash  as
shown below:

                                BA307E-SS        BA327E-SS

      

Underrange

-9.9.9.9

-9.9.9.9.9

Overrange

 9.9.9.9

 9.9.9.9.9

Although  not  guaranteed,  most  indicators  will
continue to function normally with an input current
between  1.8mA and  4mA,  at  lower  currents  the
instrument  will  display  ‘LPLo’  before  it  stops
functioning.

Under or overrange of the BA327E-SS bargraph is
indicated by an activated arrow at the appropriate
end of the bargraph and a flashing bargraph scale.

15

Summary of Contents for BA307E-SS

Page 1: ...Issue 3 4th May 2017 BA307E SS BA327E SS intrinsically safe loop powered rugged panel mounting indicators Issue 3 ...

Page 2: ...7 2 Lineariser calibration using internal reference 7 2 1 Example adding break points to a new indicator 7 3 Under and over range 7 4 Lineariser under and over range 7 5 Lineariser default configuration 8 Maintenance 8 1 Fault finding during commissioning 8 2 Fault finding after commissioning 8 3 Servicing 8 4 Routine maintenance 8 5 Guarantee 8 6 Customer comments 9 Accessories 9 1 Scale card 9 2...

Page 3: ...nd confirmed that the front of both instruments comply with Ex e Ex p and Ex t ingress and impact requirements The intrinsic safety EC Type Examination certificate specifies that under fault conditions the output voltage current and power at the 4 20mA input terminals will not exceed those specified for simple apparatus in Clause 5 7 of EN 60079 11 which simplifies installation and documentation F...

Page 4: ...x gas and dust certification This section of the instruction manual describes ATEX gas certification ATEX dust and IECEx approvals are described in Appendixes 1 and 2 3 1 ATEX gas certification Both models have been issued with common EC Type Examination Certificate ITS14ATEX28077X by Notified Body Intertek Testing and Certification Ltd This confirms compliance with harmonised European standards a...

Page 5: ...e and inductance between the two 4 20mA input terminals 1 and 3 is Ci 13nF Li 8µH Effectively 0 The maximum permitted loop cable parameters can be calculated by adding these figures to Ci and Li of other instruments in the loop and subtracting the totals from the maximum cable capacitance Co and cable inductance Lo permitted for the Zener barrier or galvanic isolator powering the loop Although the...

Page 6: ... 2 wire transmitter powered by a Zener barrier and alternatively by a galvanic isolator Fig 2a Loop powered by a Zener barrier Fig 2b Loop powered by a galvanic isolator 4 2 Remote indication The indicators may be driven via an intrinsically safe interface from a 4 20mA safe area signal to provide a remote display within a hazardous area The type of intrinsically safe interface is not critical eit...

Page 7: ...and impact protection as the front of both indicators comply with Ex e impact and ingress requirements Although mounted in an Ex e panel enclosure the BA307E SS or BA327E SS remains Group II Category 1G Ex ia IIC T5 Ga intrinsically safe apparatus so must be powered via a Zener barrier or galvanic isolator as described in section 4 1 and 4 2 of this manual Some Zener barriers and galvanic isolator...

Page 8: ...bed in section 4 1 and 4 2 of this manual to ensure that the instrument s front panel push button switches are nonincendive When installed in an Ex p panel enclosure the four vents at the rear of indicator which are shown in Fig 8 should not be obstructed Zener barriers and galvanic isolators may be installed in the same Ex p enclosure as the indicator All may be mounted in an Ex px enclosure inst...

Page 9: ...the edges of aperture are de burred b Inspect the indicator s captive gasket and ensure that it is not damaged before inserting the indicator into the panel enclosure aperture c If the enclosure panel is less than 1 0mm thick or is non metallic an optional BEKA stainless steel support plate should be slid over the rear of the indicator before the panel clamps are fitted to evenly distribute the cl...

Page 10: ...requested units of measurement if this information is not supplied when the indicator is ordered a blank card will be fitted A pack of self adhesive scale cards printed with common units of measurement is available as an accessory from BEKA associates Custom printed scale cards can also be supplied To change a scale card unclip the protruding end of the flexible strip by gently pushing it upwards ...

Page 11: ...0 Function FunC Std Std Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph start BarLo 0 00 Bargraph finish BarHi 100 00 P button in display mode C P Tare tArE Off Off Default configuration can easily be changed on site 6 1 Summary of configuration functions This section summarises each of the main configuration functions and includes a cross re...

Page 12: ...12 ...

Page 13: ...e lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or press the or button to change the setting followed by the E button to return to the configuration menu Std Linear Provides a linear relationship between the 4 20mA indicator input current and the indicator display root Square...

Page 14: ...mA If these conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the instrument will be reverse acting i e an increasing input current will cause the display to decrease 6 6 Calibration using internal reference SEt Using the SEt function the indicator can be calibrated without the need for an accurate externa...

Page 15: ...tting between on and oFF When set as required press E to return to the tARE prompt in the configuration menu 6 10 Security code CodE Access to the instrument configuration menu may be protected by a four digit security code which must be entered to gain access New instruments are configured with the default security code 0000 which allows unrestricted access to all configuration functions To enter...

Page 16: ...irect calibration of the lineariser with an external current source and is the preferred method when traceability is required If the exact system non linearity is unknown this method also allows direct calibration from the variable to be displayed e g the output from a level sensor in an irregular tank may be displayed in linear volumetric units by filling the tank with known incremental volumes a...

Page 17: ...17 ...

Page 18: ...lowing break point 7 1 1 Example Adding break points to a new indicator When adding break points to a new indicator or to a lineariser following resetting to the factory defaults using the LtAb function described in section 6 11 each additional break point should be added before the highest existing breakpoint The first additional break point should be added before the default break point 1 1 whic...

Page 19: ...he corresponding indicator display at each of the break points can now be defined using the diSP sub function Using the and buttons select the diSP sub function and press P which will reveal the starting point for the first break point 0 n where n is the total number of break points entered Press P and use the and buttons and the P button to move between digits to enter the required indicator disp...

Page 20: ...el enclosure additional precautions may be required If an indicator fails after it has been functioning correctly follow the procedure shown in section 8 1 If this does not reveal the cause of the fault it is recommended that the instrument is replaced This can be done without disconnecting power but while the indicator is disconnected the 4 20mA loop will be open circuit 8 3 Servicing All BA307E ...

Page 21: ...rent powering the indicator is removed both alarm outputs will open irrespective of configuration The open circuit condition should therefore be chosen as the alarm condition when designing an alarm system Fig 13 illustrates the conditions available and shows which are fail safe When an alarm occurs an alarm annunciator on the indicator front panel is activated and if required the numerical displa...

Page 22: ...22 ...

Page 23: ...l facilities Summary of alarm configuration functions Display Description of function EnbL Alarm enable Enables or disables the alarm without changing the alarm parameters See section 9 3 4 SP1 Alarm setpoint 1 Adjusts the alarm setpoint The alarm is activated when the indicator display equals the setpoint See section 9 3 5 Hi Lo Alarm function Defines the alarm function as High or Low See section...

Page 24: ... fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press P to reveal the setting The function may be changed by pressing the or button followed by the E button to return to the alarm configuration menu 9 3 8 Hysteresis HStr Hysteresis is shown in the units t...

Page 25: ...ng access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct setpoint access menu is enabled and the separate security code entered from the ACSP function in the alarm configuration menu as shown in Fig 15 To change the menu parameters select ACSP from the config...

Page 26: ...layed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 18 Fig 18 Displayed value and setpoints on bargraph The bargraph area below the low alarm setpoint SP1 and the bargraph area above the high alarm setpoint SP2 are activated The displayed variable is represented by an activated single bar which moves between the low ...

Page 27: ...xceed those specified by clause 5 7 of EN 60079 11 for simple apparatus which simplifies system design and documentation Providing the increased voltage drop can be tolerated the intrinsic safety and system design described in sections 3 and 4 of this manual remain valid with the loop powered backlight 9 4 2 Separately powering the backlight The optional backlight may also be powered from a separa...

Page 28: ... is likely to occur occasionally in normal operation Zone 22 explosive atmosphere in the form of a cloud of combustible dust in air is not likely to occur in normal operation but if it does occur will only persist for a short period Be used with dust in subdivisions IIIA combustible flyings IIIB non conductive dust IIIC conductive dust For use with IIIC conductive dusts special conditions for safe...

Page 29: ...he inside of an Ex t enclosure is effectively a safe area and the certified intrinsically safe interface may therefore be mounted in the same Ex t panel enclosure as the indicator The indicator terminals the wiring to the indicator and the intrinsically safe interface if mounted within the enclosure should be segregated from all other non intrinsically safe wiring and equipment within the panel en...

Page 30: ...of this manual may therefore be used for IECEx installations but the local code of practice should also be consulted A2 3 Special conditions for use in IIIC dusts The IECEx intrinsic safety certificate has an X suffix indicating that for some applications special conditions apply for safe use which are identical to ATEX special conditions for safe use a When installed in an Ex px Ex py or Ex pz pa...

Page 31: ... parameters are similar to the ATEX and IECEx parameters and the systems shown in section 4 of this manual may therefore also be used for FM and cFM installations The FM and cFM Control Drawing CI300 72 for intrinsically safe applications which is contained in this appendix specifies that when correctly installed a BA307E SS or a BA327E SS will not invalidate the certification of an AEx e AEx p or...

Page 32: ...32 ...

Page 33: ...33 ...

Page 34: ...34 ...

Page 35: ...35 ...

Page 36: ...36 ...

Page 37: ...37 ...

Page 38: ...38 ...

Page 39: ...39 ...

Page 40: ...40 ...

Page 41: ...41 ...

Page 42: ...42 ...

Reviews: