background image

13

6.4  Position of the decimal point:   ‘dP’

A dummy decimal point can be positioned between
any of the digits or it may be absent.  To position
the decimal point select 'dP' from the menu and
press  

P

.      The  decimal  point  can  be  moved  by

pressing the 

 or 

 push button.  If a decimal

point is not required it should be positioned beyond
the  most  or  least  significant  digit.      When
positioned as required press the 

E

 button to enter

the selection and return to the configuration menu.

6.5  Calibration using an external
        current source:   ‘CAL’

This  function  enables  the  zero  and  span  of  the
indicator  to  be  adjusted  using  an  external
calibrated  current  source.    When  used  with  an
accurate  traceable  current  source  this  is  the
preferred method of calibration.

Zero is the indicator display with 4mA input
Span is the indicator display with 20mA input

To  calibrate  the  indicator  select  'CAL'  from  the
configuration menu and press 

P

.

  The indicator will

display 'ZEro'  which is a request for a 4mA input
current.    Adjust  the  external  current  calibrator  to
4.000mA and again press  

P

 

which will reveal the

current  zero  display.      The  flashing  digit  of  the
indicator display can be changed by pressing the

 or 

 buttons, when set as required pressing 

P

will transfer control to the next digit.  When all the
digits  have  been  adjusted,  press  

E

  to  enter  the

new zero and return to the 'ZEro' prompt .

Pressing the 

 button will cause the indicator to

display 'SPAn' which is a request for a 20mA input
current.    Adjust  the  external  current  calibrator  to
20.000mA and again press 

P

 

which will reveal the

existing  span  display.      The  flashing  digit  of  the
indicator display can be changed by pressing the

 or 

 buttons, when set s required pressing  

P

will transfer control to the next digit.  When all the
digits have been adjusted press 

E

 t

o enter the new

span  and  return  to  the  'SPAn'  prompt.    Finally
press 

E

 

again to return to the configuration menu.

Notes:

a. The indicator input current must be adjusted to

the   required   value   before   the   zero   and   span
functions are entered by pressing the 

 button.

b. Indicators may be calibrated at currents other

than   4   and   20mA,   withiin   the   range   3.8   to
21.0mA providing the difference between the
two  currents  is  greater  than  4mA.    If  these
conditions are not complied with, the indicator
displays ‘FaiL’ and aborts the calibration.

 

c. If  the  zero  current  is  greater  than  the  span

current  the  instrument  will  be  reverse  acting
i.e. an increasing input current will cause the
display to decrease.

6.6 Calibration using internal reference:    

   ‘SEt’

Using  the  ‘SEt’  function  the  indicator  can  be
calibrated without the need to know the value of
the  4/20mA  input  current,  or  to  disconnect  the
indicator from the 4/20mA loop.  

When  using  the  ‘Set’  function  the  indicator’s
internal reference is used to simulate a 4mA and
20mA input current.  

Zero is the display with a simulated 4mA input
Span is the display with a simulated 20mA input

To calibrate the indicator display select 'SEt' from
the configuration menu and press

 

P

.  

The indicator

will display 'ZEro', pressing 

P

 again will reveal the

current display at 4mA.   The flashing digit can be
adjusted by pressing the 

 or 

 buttons,   when

the flashing digit is correct pressing 

P

 

will transfer

control to the next digit.  When all the digits have
been  adjusted,  press  

E

  to  return  to  the  'ZEro'

prompt.

To adjust the display at 20mA, press the 

 button

which  will  cause  the  indicator  to  display  'SPAn',
pressing  

P

 

will then reveal the indicator’s existing

display  at  20mA.      The  flashing  digit  can  be
adjusted by pressing the 

 or 

 buttons, when

the flashing digit is correct pressing 

P

 will transfer

control to the next digit.  When all the digits have
been  adjusted  press  

E

  t

o  return  to  the  'SPAn'

prompt followed by 

E

 to return to the ‘SEt’ prompt

in the configuration menu.

6.7  Bargraph format and calibration:   ‘bAr’
       

Only the BA327NE has a bargraph

In  addition  to  a  five  digit  numerical  display  the
BA327NE  has  a  31  segment  analogue bargraph
which  may  be  configured  to  start  and  finish
anywhere within the indicator’s numerical display
range.

To  configure  the  bargraph  select  'bAr'  from  the
configuration menu and press

 

P

.  

The indicator will

display  'tYPE',  pressing  

P

  again  will  reveal  the

existing  bargraph  justification  which  can  be
changed to one of the following four or five options
using the 

 or 

 button:

                       

Bargraph justification starts from

‘LEFt’

Left end of display

‘CEntr’

Centre of display

‘riGHT’

Right end of display

‘AlrSP’

Only with alarms – see section 9.3

‘oFF’

Bargraph disabled

When  set  as  required  press  

E

  to  return  to  the

‘tYPE’ sub-function prompt.

Summary of Contents for BA307NE

Page 1: ...Issue 3 1st Novemeber 2017 BA307NE BA327NE Ex nA Ex tc loop powered panel mounting indicators Issue 3...

Page 2: ...figuration 8 Maintenance 8 1 Fault finding during commissioning 8 2 Fault finding after commissioning 8 3 Servicing 8 4 Routine maintenance 8 5 Guarantee 8 6 Customer comments 9 Accessories 9 1 Scale...

Page 3: ...European ATEX Directive 94 9 EC The certificate number has an X suffix indicating that the two indicators are subject to special condition for safe use For installations in Zone 2 the panel enclosure...

Page 4: ...both instruments with a common Type Examination Certificate number ITS14ATEX48028X This has been used to confirm compliance with the European ATEX Directive for Group II Category 3GD equipment The in...

Page 5: ...roups Group A propane Group B ethylene Group C hydrogen In gases that may safely be used with equipment having a temperature classification of T1 450o C T2 300o C T3 200o C T4 135o C T5 100o C At ambi...

Page 6: ...ea should be individually fused and contain a means of isolation These two requirements are frequently satisfied by using switch fuse holders with easily removable fuses which can be extracted to achi...

Page 7: ...is a satisfactory method at the low voltages and currents common in instrumentation systems Clear identification of and easy access to the means of isolation is essential for their effective use It i...

Page 8: ...el enclosure cut out dimensions Fig 4 Dimensions 5 2 Installation Procedure a Cut the aperture specified in Fig 4 in the panel enclosure Ensure that the edges of aperture are de burred b Inspect the i...

Page 9: ...ard showing the requested units of measurement if this information is not supplied when the indicator is ordered a blank card will be fitted A pack of self adhesive scale cards printed with common uni...

Page 10: ...10...

Page 11: ...uration BA307NE BA327NE Access code CodE 0000 0000 Function FunC Linear Linear Display at 4mA Zero 0 0 0 00 Display at 20mA SPAn 100 0 100 00 Resolution rESn 1 digit 1 digit Bargraph start BarLo 0 00...

Page 12: ...in 16 segment adjustable lineariser To reveal the existing indicator function select FunC from the configuration menu and press P If the function is set as required press E to return to the menu or pr...

Page 13: ...tween the two currents is greater than 4mA If these conditions are not complied with the indicator displays FaiL and aborts the calibration c If the zero current is greater than the span current the i...

Page 14: ...rompt in the configuration menu 6 10 Security code CodE Access to the instrument configuration menu may be protected by a four digit security code which must be entered to gain access New instruments...

Page 15: ...e This method allows direct calibration of the lineariser with an external current source and is the preferred method when traceability is required If the exact system non linearity is unknown this me...

Page 16: ...16...

Page 17: ...to the required value before the P button is operated to enter the required indicator display Error message If during calibration the indicator displays a FAiL error message the current at which the s...

Page 18: ...oints has been defined and then return to the in sub function by pressing the E button The corresponding indicator display at each of the break points can now be defined using the diSP sub function Us...

Page 19: ...rior of the instrument with dust or moisture Some end users may prefer not to permit live maintenance to minimise risk ENSURE PLANT SAFETY BEFORE STARTING MAINTENANCE Live maintenance within the hazar...

Page 20: ...the enclosure panel being distorted or creeping when the indicator panel clamps are tightened and ensures that the joint between the indicator and the panel enclosure in which it is mounted remains se...

Page 21: ...ed in Zone 2 is displaying the output from an Flameproof Ex d 2 wire transmitter located in Zone 1 Alarm 1 is switching an Ex e sounder in Zone 1 and alarm 2 is switching a sounder located in the safe...

Page 22: ...22...

Page 23: ...ndicator display equals the setpoint See section 9 4 5 Hi Lo Alarm function Defines the alarm function as High or Low See section 9 4 6 no nC Normally open or normally closed output Sets the alarm out...

Page 24: ...r fail safe operation both alarm outputs should be conditioned to be open in the alarm condition nC To check or change the alarm output status select no nC from the alarm configuration menu and press...

Page 25: ...ving access to the indicator configuration menu Protection against accidental adjustment of the setpoints when the indicator is in the display mode is provided by a separate security code This direct...

Page 26: ...int plus the displayed value to be represented or a low and a high setpoint plus the displayed value to be represented by the bargraph as shown in Fig 16 Fig 16 Displayed value and setpoints on bargra...

Page 27: ...o or greater than 9V At lower supply voltages the backlight will continue to function but with reduced brilliance and will draw less current The ATEX certificate therefore only specifies the maximum s...

Page 28: ...re four design requirements 1 The indicator should be installed generally as shown in section 5 and Fig 20 2 The certificate specifies that the indicator should be connected to a limited energy 4 20mA...

Page 29: ...Fig 21 2 The four vents at the rear of the indicator enclosure shown in Fig 22 must not be obstructed 3 The certificate specifies that the indicator should be connected to a limited energy 4 20mA circ...

Page 30: ...ne 22 The panel enclosure containing the indicator should be located where the minimum amount of dust will accumulate on them and the installation should be as specified in section 5 of this instructi...

Page 31: ...X 4 ETL cETL certification for installations in USA and Canada A4 0 cETL Mark For installations in the USA and Canada the BA307NE and BA327NE indicators have ETL and cETL Ex nA and Ex tc approval Cont...

Reviews: