background image

9

Display

Description of function

'rEF'

Calibration of internal references

The indicator contains two  references
representing a 4 and 20mA. input
These enable the BA324ND display to
be calibrated without the need for an
external calibrator, or disconnection
from the 4/20mA loop. See the SEt
function.
If the SEt or C--P functions are to be
used, the internal references should
be periodically calibrated.
See section 6.7.2

'C - - P' Function of P push-button

The BA324ND  may be programmed
to display the input current in
milliamps or the input current as a
percentage when the 

P

  push-button

is operated.
See section 6.8

'COdE'

Security code

Defines a four digit numeric code
which must be entered to gain access
to programmable functions.  Default
code 0000 disables the security
function and allows unrestricted
access to all programmable functions.
See section 6.9

6.2 Root extractor:   root

This function is primarily intended for use with
differential flowmeters which have a square law
4/20mA output.  To activate the square root
extractor select 'root' from the menu and press 

P

which will reveal if the function is 'On' or 'OFF'.   If
the function is set as required, press 

E

 to return to

the menu, or press the 

Up

 or 

Down

 button to

change the setting, followed by the 

E

 button to

return to the main menu.  For reference, the
following table shows the output current from a
non-linearised differential flowmeter.  Below 5% of
flow the BA324ND display is forced to zero.

% of full flow

Current output mA

5.0

4.04

10.0

4.16

25.0

5.00

50.0

8.00

75.0

13.00

100.0

20.00

6.3 Resolution:   rESn

This function defines the resolution of  the least
significant display digit.  Decreasing the display
resolution can improve the readability of a noisy
signal.   Select 'rESn' from the menu and press 

P

which will reveal the current display resolution.  To
change the resolution press the 

Up

 or 

Down

 button

to select 1, 2, 5 or 10 digits, followed by the 

E

button to return to the menu.

6.4 Position of the decimal point:   d.P.

A dummy decimal point can be positioned between
any of  the digits or may be absent.  To position
the decimal point select 'd.P.' from the menu and
press 

P

.  The decimal point can then be moved or

turned off by pressing the 

Up

 or 

Down

 push-

button, followed by 

E

 to return to the menu.

6.5 Calibration of display using an external

current source:   CAL

This function enables the zero and span of the
indicator to be adjusted using an external
calibrated current source.  When used with an
accurate traceable current source this is the
preferred method of calibration.

To calibrate the indicator select 'CAL' from the
main menu and press 

P

.  The indicator will display

'ZErO'  which is a request for a 4mA input current.
Adjust the external current calibrator to 4.000mA
and again press 

P

  which will reveal the current

zero display.   Each digit of the indicator display
can be changed by pressing the 

Up

 or 

Down

buttons. When the first digit is correct pressing P
will transfer control to the next digit.

Summary of Contents for BA324ND

Page 1: ...Issue 7 4th March 2014 BA324ND Type nL certified loop powered 4 digit field mounting indicator Issue 7 Only for replacement use after May 2013 Use BA324NE for new applications...

Page 2: ...eferences 8 Maintenance 8 1 Fault finding during commissioning 8 2 Fault finding after commissioning 8 3 Servicing 8 4 Routine maintenance 8 5 Guarantee 8 6 Customer comments 9 Accessories 9 1 Units o...

Page 3: ...forward biased diode D1 The voltage developed across D1 which is relatively constant is multiplied by a switch mode power supply and used to power the instrument The voltage developed across R1 which...

Page 4: ...60079 Part 14 Electrical Installation in Hazardous Areas When designing systems for installation outside the UK the local Code of Practice should be consulted 3 3 4 20mA input terminals Input terminal...

Page 5: ...Fig 2 BA324ND in a transmitter loop In practice it is only necessary to ensure that in normal operation the maximum current flowing in the loop is less than 30mA In the example shown in Fig 2 this cu...

Page 6: ...ted to the internal EMC filters For maximum radio frequency interference rejection this terminal should be connected to a local earth or to a cable screen which is earthed in the safe area To prevent...

Page 7: ...and may be operated when a flammable atmosphere is likely to be present All the programming functions are contained in an easy to use menu which is shown diagramatically in Fig 6 Each function is sum...

Page 8: ...s the resolution of the least significant display digit May be set to 1 2 5 or 10 digits See section 6 3 d P Decimal point Positions the dummy decimal point between any of the digits or turns it off S...

Page 9: ...ial flowmeter Below 5 of flow the BA324ND display is forced to zero of full flow Current output mA 5 0 4 04 10 0 4 16 25 0 5 00 50 0 8 00 75 0 13 00 100 0 20 00 6 3 Resolution rESn This function defin...

Page 10: ...10...

Page 11: ...least significant digit has been adjusted press E to return to the SPAn prompt followed by E to return to the menu 6 7 Conditioning sub menu Cond This sub menu allows the mains line frequency at whic...

Page 12: ...ith the default security code 0000 which allows unrestricted access to all programming functions To enter a new security code select COdE from the menu and press P which will cause the indicator to di...

Page 13: ...and press E to return to the main menu Step 4 Select frequency of maximum mains line rejection Scroll though the main menu until Cond is displayed Enter the sub menu by pressing P twice and select the...

Page 14: ...y of maximum mains line rejection Scroll though the main menu until Cond is displayed Enter the sub menu by pressing P twice and select the FrE function Using the Up or Down buttons select 50 and then...

Page 15: ...ay has a large voltage ripple content Unable to Incorrect Enter correct security enter the security code or fit security programme code link in override position Mode entered See Fig 7 8 2 Fault findi...

Page 16: ...intenance The mechanical condition of the instrument and electrical calibration should be regularly checked The interval between inspections depends upon environmental conditions We recommend that ini...

Page 17: ...A324ND can also be supplied with a blank or custom engraved stainless steel plate secured to the instrument terminal cover by two screws 9 2 Alarms The BA324ND can be supplied with two solid state sin...

Page 18: ...owing table summarises each of the alarm programme functions and includes a cross reference to more detailed information Again only the functions on alarm 1 are listed but alarm 2 has identical facili...

Page 19: ...een calibrated to display 0 with 4mA input and 10000 with 20mA input the two alarm setpoints may be positioned anywhere between 125 and 10125 To adjust the setpoint select SP1 or SP2 from the alarm me...

Page 20: ...can be changed by pressing the Up or Down button followed by the E button to return to the alarm menu 9 2 8 Hysteresis HStr During programming hysteresis is shown in the units the indicator has been c...

Page 21: ...nt AcSP This function controls a separate menu which provides direct access to the alarm setpoints when the indicator is in the display mode See section 9 2 12 for a full description An operator may t...

Page 22: ...ressing E will return the display to the SP1 or SP2 prompt from which the other setpoint may be selected or the indicator may be returned to the display mode by pressing E again Direct access to the a...

Page 23: ...n can be selected using the Up and Down push buttons The delete break point function dEL operates in a similar manner to the Add function described above When the required number of break points has b...

Page 24: ...24...

Page 25: ...equired break point has been selected press P to reveal the existing display Using the Up and Down buttons and P to move between digits set the revised display Press E to return to the sub menu from w...

Page 26: ...i 30V dc Ii 100mA dc Pi 1 3W This allows the backlight to be powered from a safe area 18 to 30V dc supply as shown in Fig 15 Providing the supply voltage is less than 30V the backlight safety componen...

Page 27: ...re available for securing the BA324ND to a horizontal or vertical pipe BA392D Stainless steel bracket secured by two worm drive hose clips Will clamp to any pipe with outside diameter between 60 and 8...

Reviews: