29
9.4
Display backlight
The BA326C can be supplied with LED
backlighting to improve display contrast when
the indicator is installed in poorly illuminated
areas. The backlight is segregated from the
measuring circuit and has been certified as a
separate intrinsically safe circuit
.
The backlight must be powered from the safe
area via a Zener barrier or a galvanic isolator
as shown in Fig 18.
Any certified Zener
barrier may be used, providing the output
parameters do not exceed:
Uo
28V dc
Io
159mA
Po
0.8W
The maximum equivalent capacitance and
inductance of the display backlight is:
Ci
=
30nF
Li
=
10µH
To determine the maximum permissible cable
parameters,
the
equivalent
backlight
parameters must be subtracted from the
maximum cable capacitance and inductance
specified by the certificate of the Zener barrier
or galvanic isolator powering the backlight.
Fig 18 Backlight power supply
The display brilliance depends upon the
current flowing through the backlight. This is
determined by the supply voltage and the end-
to-end resistance of the Zener barrier or output
resistance of the galvanic isolator. Brilliance
will not be significantly reduced until the
current falls below 20mA.
Backlight current = Vsupply –18
End-to-end resistance of
barrier
#
# or output resistance of galvanic isolator