background image

26

9.2.6  Setpoint adjustment:   ‘SPx1’ and
                                                 ‘SPx2’

The rate alarm setpoints SPr1 and SPr2 may
be positioned anywhere between 0 and
999999, and the total alarm setpoint SPt1 and
SPt2 anywhere between 00000000 and
99999999.

To adjust the setpoint of alarm 1 select 'SPr1'
or ‘SPt1’, depending upon how it has been
programmed, from the alarm menu. Press 

P

which will reveal the existing value with one
digit flashing.  Each digit of the setpoint can be
adjusted using the 

Up

 and 

Down

 push-buttons,

and the 

P

 button to move to the next digit.  The

position of the decimal point is defined by the
d.P. function which should be set first.  When
the required value has been entered press 

E

to return to the alarm menu.

Note:

 ‘SPr1’ is displayed when alarm 1 has

been programmed as a rate alarm and ‘SPt1’
when programmed as a total alarm.

9.2.7  Alarm function:   ‘HI.LO’

Each alarm must be conditioned as a high or
low alarm.   To check or change the alarm
function select 'HI.LO' from the alarm menu
and press 

P

 to reveal the current setting.  The

function can be changed by pressing the 

Up

 or

Down

 buttons followed by the 

E

 button to

return to the alarm menu.

9.2.8  Alarm output status:   ‘no.nc’

The two alarm outputs may be open or closed
in the non-alarm condition.  When the BA344D
power supply is turned off or disconnected,
both the alarm outputs will open irrespective of
whether normally open or normally closed
outputs have been selected.  Therefore, when
designing an alarm system, normally closed
‘nc’ should be selected so that the output
opens when an alarrm occurs or if the power
supply fails.

To check the alarm output status select 'no.nc'
from the alarm menu and press 

P

 to reveal the

current setting.  The function can be changed
by pressing the 

Up

 or 

Down

 button followed by

the 

E

 button to return to the alarm menu.

9.2.9  Hysteresis:   ‘HStr’

Hysteresis is only available on rate alarms.
During programming hysteresis is shown in the
units of rate the BA344D has been calibrated
to display.

To adjust the hysteresis, select 'HStr' from the
alarm menu and press 

P

 which will reveal the

existing figure.  Each digit can be adjusted
using the 

Up

 and 

Down

 push-buttons, and the

P

 button to move to the next digit.  When the

required hysteresis has been entered, press 

E

to return to the alarm menu.

e.g.  A BA344D calibrated to display a flow of 0
to 5000, with a high alarm set at 4000 and
hysteresis of 100 will perform as follows:

High alarm will be activated when flow
equals or exceeds 4000, but will not reset
until the flow falls below 3900.

9.2.10  Alarm delay:   ‘dELA’

This function enables activation of the alarm
output to be delayed for a fixed time following
the alarm condition being detected.  The delay
can be programmed in 1 second increments up
to 3600 seconds.  If a delay is not required
zero should be entered.  To adjust the delay
select 'dELA' from the alarm menu and press 

P

which will reveal the existing delay time.  Each
digit of the delay can be adjusted using the 

Up

and 

Down

 push-buttons, and the 

P

  button to

move to the next digit.  When the required
delay has been entered, press 

to return to

the alarm menu.

The alarm annunciator will start flashing
immediately an alarm is detected and continue
for the delay time after which the alarm output
will be activated.

9.2.11  Alarm silence time:   ‘SIL’

The alarm silence function is primarily intended
for use in small installations where the alarm
output directly operates an annunciator such
as a sounder.  When the alarm silence time is
set to any figure other than zero, the 

P

 push-

button becomes an alarm accept button.

After an alarm has occurred, operating the 

P

button will cause the alarm output to revert to
the non-alarm condition for the programmed
alarm silence time.  When an alarm is silenced
the alarm annunciator will flash until the silence
time expires.

To adjust the silence time select 'SIL' from the
alarm menu and press 

P

  which will reveal the

existing time.  The time may be adjusted to
between 0 and 3600 seconds in 1 second
increments.  Each digit can be adjusted using
the 

Up

 and 

Down

 push-buttons, and the 

P

button transfers control to the next digit.  When
the required time has been entered press 

to

return to the alarm menu.

Summary of Contents for BA344D

Page 1: ...Issue 5 7th April 2009 BA344D Intrinsically safe Battery powered pulse input field mounting rate totaliser Issue 5...

Page 2: ...0 Rate filter 6 11 Total scale factor 6 12 Clip off 6 13 Local reset 6 14 Resetting grand total 6 15 Security code 7 Calibration Example 7 1 Calibration procedure 8 Maintenance 8 1 Fault finding durin...

Page 3: ...es or via the optional external keypad Optional alarms which may be programmed to function on the rate or total displays further extend the applications of this instrument The BA344D contains an inter...

Page 4: ...mospheres which conform with BS EN60079 Part14 2003 Electrical Installation in Hazardous Areas When designing systems for installation outside the UK the local Code of Practice should be consulted For...

Page 5: ...and 4 is Ci 30nF Li 10 H To determine the maximum permissible cable parameters these figures should be subtracted from the maximum permitted cable parameters specified for the device connected to ter...

Page 6: ...perated simultaneously See 6 13 Each time the total display is reset to zero the grand total is transferred to permanent memory 3 5 Zones gas groups and T rating The BA344D has been issued with an EC...

Page 7: ...onnected to terminals 5 6 The maximum input counting frequency is 10kHz 4 1 System certificates When the input signal is derived from the safe area or the total display is to be reset from within the...

Page 8: ...rconnected within the BA344D see Fig 1 4 3 Use with Galvanic Isolators Galvanic isolators are probably the simplest intrinsically safe interface as they provide isolation and do not require a high int...

Page 9: ...nstallation procedure a Remove the instrument terminal cover by unscrewing the two captive A screws b Mount the instrument on a flat surface and secure with two M6 screws through the B holes Alternati...

Page 10: ...ssembly The discharged battery may be removed and the replacement fitted while the BA344D rate totalisers is installed in the hazardous area Fig 7 illustrates the replacement sequence a If the instrum...

Page 11: ...the instrument enclosure Before replacing the terminal cover turn the instrument ON by moving the slide switch to the right Fig 7 Replacement of internal battery 5 5 Disposal of used batteries The BEK...

Page 12: ...ond minute or hour Similarly to calculate the total flow in the required units of measurement the number of input pulses is divided by the total scale factor SCALE t Total flow is continuously added t...

Page 13: ...13...

Page 14: ...section 6 6 dISPLAY 2 Lower display Turns the lower display which normally shows rate on or off See section 6 7 Display Description of function SCALE r Rate Scale Factor Defines the arithmetic relatio...

Page 15: ...ype select InPut from the menu and press P which will reveal the current input type Pressing Up or Down will scroll through the four options UOLts Voltage pulse input Threshold 1V and 3V 10kHz max Inp...

Page 16: ...int position and simultaneously press the Up and Down buttons When the required rate scale factor has been entered press E to store the number and return to the main menu 6 9 Timebase t bASE The timeb...

Page 17: ...ecimal point in the scale factor move the flashing digit to the left hand side of the required decimal point position and simultaneously press the Up and Down buttons When the required total scale fac...

Page 18: ...d before access to the programme menu is granted New instruments are programmed with the default security code 0000 which allows unrestricted access to all programming functions without entering a sec...

Page 19: ...3 Select the type of input Using the Up or Down button scroll through the main menu until InPut is displayed then press P to reveal the existing setting Select COIL the input for a magnetic pick off u...

Page 20: ...A344D is required to display 1 cubic metre for every 105 x 1 000 pulses received SCALE t should therefore be set to 105 000 Select SCALE t from the main menu and press P which will reveal the current...

Page 21: ...rument to the CLr Gtot prompt in the main menu Step 12 Define the security code Defining a security code prevents unauthorised access to the programme functions Select COdE from the main menu and pres...

Page 22: ...equipment should be used unless a gas clearance certificate is available If a BA344D fails after it has been functioning correctly the following table may help Symptom Cause Check No display Instrume...

Page 23: ...larm annunciators on the instrument display indicate the status of both alarms If an alarm delay or silence time has been selected the annunciator will flash during the delay or silence period Program...

Page 24: ...lay Description of function EnbL Alarm enable Enables or disables the alarm function without changing the alarm parameters See section 9 2 4 tYPE Type of alarm Defines whether the alarm operates on th...

Page 25: ...rn to the alarm menu 9 2 5 Type of alarm tYPE Alarm 1 and Alarm 2 may both be rate or total alarms or one may be conditioned for rate and the other for total To check or change the type of alarm selec...

Page 26: ...select HStr from the alarm menu and press P which will reveal the existing figure Each digit can be adjusted using the Up and Down push buttons and the P button to move to the next digit When the req...

Page 27: ...o the setpoints in the display mode by pressing the P and Up buttons simultaneously New instruments with alarms are supplied with the security code set to 0000 9 2 13 Adjusting alarm setpoints from th...

Page 28: ...mounting kits are available for securing the BA344D to a horizontal or vertical pipe BA392D Stainless steel bracket secured by two worm drive hose clips for 60 to 80mm outside diameter pipes BA393 He...

Page 29: ...ertification Appendix 1 EnbL 9 2 3 9 2 4 Fault finding during commissioning 8 1 after commissioning 8 2 FILtEr 6 10 Flow indicator 2 2 Galvanic isolators 4 3 Gas groups 3 5 Grand total 2 2 clear 6 14...

Page 30: ...ve atmosphere in the form of a cloud of combustible dust in air is likely to occur occasionally in normal operation Zone 22 explosive atmosphere in the form of a cloud of combustible dust in air is no...

Page 31: ...ctioning immediately it is switched on When a replacement battery is fitted or when the rate totaliser is turned on after not being used for an extended period the battery should be conditioned by dra...

Reviews: