Press 

P

  to show the transmitter display at which

the output current is 20mA.  The flashing digit of the
display can be changed by pressing the 

Up

 or

Down

 buttons.  When the first digit is correct,

pressing 

P

 will transfer control to the next digit.

When the least significant digit has been adjusted,
press 

E

 to return to the 'SPAn' prompt.

Note:

 If a complete loop including a primary

element, BA378C temperature transmitter and a
load are being calibrated, the Trim Menu is the
preferred method. It allows the transmitter display
and output current to be adjusted to compensate
for inaccuracies in the other components in the
loop.  Details of the trim menu are contained in
section 7 which also describes periodic calibration.

6.7  Security Code  'COdE'

Access to the Programme and Trim Menus may be
protected by a four digit numerical security code
which must be entered before access is granted.
New instruments are programmed with the default
security code 0000 which allows unrestricted
access to both menus.

To enter a new security code select 'COdE' from
the Programme Menu and press 

P

 which will cause

the transmitter to display the current security code.
Each digit of the code can be changed using the 

Up

and 

Down

 push-buttons, and the

  P

 button to move

to the next digit.  When the required code has been
entered press 

E

 to return to the main menu.  The

revised security code will be activated when the
transmitter is returned to the display mode.

If the security code is lost, access to the
programme and Trim Menus can be obtained by
putting the internal security link in the override
position - see Fig 8.  The security code can then be
viewed by selecting 'COdE' from the menu and
pressing 

P

.

To access the security code link un-plug the
removable terminal(s) from the rear of the
instrument by gently pulling, and remove the
instrument rear panel which is secured by six
screws.  The link can then be moved to the override
positioned as shown in Fig 8.

Fig 8  Location of security code override link

6.8  Calibrating the internal references

WARNING!

Changing the internal references will
alter the instrument performance. Do
not attempt to recalibrate these
references unless you have adequate
test equipment and have read the
following instructions. For routine
calibration of the instrument use the
Trim Menu described in Section 7.

If the Trim Menu is 

not

 used for routine calibration

of the BA378C transmitter, for maximum accuracy
the internal references may require periodic
checking against external standards. Annual checks
will produce acceptable performance, although less
frequent checks are usually adequate.  

If the BA378C transmitter is calibrated against an
external voltage, resistance or temperature
calibrator using the Trim Menu, the internal
references should never need re-calibrating.
Details  of  the  Trim  Menu  are  contained  in
section 7.

There are four internal references:

Voltage input
3-wire RTD
4-wire RTD
Current output  

The input voltage reference must be adjusted first
irrespective of the transmitter application, but  it is
only necessary to adjust the other three references
if the function to which they relate is being used.
For example, a BA378C with a thermocouple input
from which the 4/20mA output current is being used
would require the voltage input and the current
output reference adjusting.

      16

Summary of Contents for BA378C

Page 1: ...BA378C Intrinsically safe panel mounting indicating temperature transmitter issue 6 Issue 6 25th September 2008 ...

Page 2: ...isplay calibration 6 6 4 20mA output current calibration 6 7 Security code 6 8 Calibrating the internal references 6 8 1 Conditioning internal references 6 8 2 Voltage input reference 6 8 3 3 wire RTD reference 6 8 4 4 wire RTD reference 6 8 5 Output current reference 7 Trim Menu 7 1 Calibration procedure using Trim Menu 7 2 Periodic re calibration using Trim Menu 8 Calibration example 9 Maintenan...

Page 3: ...ions of Section 6 you will find information on setting up for Section 6 3 Thermocouple input Section 6 4 Resistance thermometer input Section 6 5 Voltage input A word of warning Section 6 8 explains how you can calibrate the instrument s internal references It is only necessary to carry out this operation if you will be using these internal references to calibrate the instrument display i e not us...

Page 4: ...4 ...

Page 5: ...lies linearisation using look up tables stored in permanent memory The resulting temperature is shown on the internal display and controls the 4 20mA output current The transmitter and the optional alarms are powered from the 4 20mA current loop but the optional display backlight requires a separate supply Each time power is applied to the instrument initialisation is performed After a short delay...

Page 6: ...gases in groups Group A propane Group B ethylene Group C hydrogen Having a temperature classification of T1 450o C T2 300o C T3 200o C T4 135o C T5 100o C This allows the BA378C to be installed in all Zones and to be used with most common industrial gases at ambient temperature between 40 and 60o C Note minimum operating temperature is 20o C 3 4 Certification information label The certification in...

Page 7: ...is earthed at the barrier busbar Fig 2 BA378C powered from floating supply If a common power supply is used to operate more than one loop the negative side of the supply is normally connected to earth To enable the negative side of the 4 20mA load also to be earthed it is necessary to have a Zener barrier in each wire going into the hazardous area as shown in Fig 3 Any certified diode return barri...

Page 8: ... the instrument and the panel The trans mitter must be positioned such that the environ mental conditions specified by the instrument datasheet and the EC Type Examination Certificate are not exceeded and it will not be attacked by aggresive substances Please consult BEKA associates if the instrument is to be installed in an area subject to high vibration Fig 5 shows the overall dimensions of the ...

Page 9: ...ribes periodic calibration When the transmitter is fitted with alarms additional functions are added to the programme menu These are described in section 11 of this manual Throughout this manual push buttons are shown in italics e g P or Up push button and legends displayed by the transmitter are shown within inverted commas e g CAL and ALr2 Access to the Programme Menu is obtained by operating th...

Page 10: ...10 ...

Page 11: ...11 ...

Page 12: ...ation Figs 7A and 7B illustrate the location of each function within the menu Display Summary of Programme Menu functions InPut Transmitter input Conditions the transmitter to accept thermocouple tHC resistance thermometer rtd or a voltage UOLt input See section 6 2 1 Display Summary of Programme Menu functions For Thermocouple Inputs tYPE Sensor type Selects the type of thermocouple input See sec...

Page 13: ...programme structure shown in Figs 7A and 7B 6 2 1 Transmitter input InPut This function conditions the transmitter to accept a thermocouple tHC resistance thermometer rtd or a voltage UOLt input To define the transmitter input select InPut from the Programme Menu and press P which will reveal the current setting This setting may be changed by scrolling through the menu using the Up or Down buttons...

Page 14: ...rrent setting The setting may be changed by pressing the Up or Down button When the required setting has been selected press E to return to the sub menu 6 4 For resistance thermometer inputs 6 4 1 Types of resistance thermometer tYPE The transmitter may be conditioned to accept 3 wire 4 wire or differential Pt100 resistance thermometers After selecting resistance thermometer input from the InPut s...

Page 15: ...y The calibration is performed using the internal references and may be made with any transmitter input voltage Select In HI from the menu and press P which will access a sub menu containing two functions InPut and dISP Select the input voltage InPut and press P which will cause the specified higher input millivoltage to be displayed in the form XX XX The flashing digit of the input voltage can be...

Page 16: ...ty link in the override position see Fig 8 The security code can then be viewed by selecting COdE from the menu and pressing P To access the security code link un plug the removable terminal s from the rear of the instrument by gently pulling and remove the instrument rear panel which is secured by six screws The link can then be moved to the override positioned as shown in Fig 8 Fig 8 Location of...

Page 17: ...dard resistor or calibrator with the compensation loop remaining connected between terminals 3 and 4 It is preferable to make this substitution as close to the RTD as possible but transmitters may be calibrated in a workshop Select 3rtd from the Cond sub menu and press P The transmitter will display scrolling decimal points while the reference is being updated followed by Ent prior to returning to...

Page 18: ...tected by a security code COdE will be displayed first Pressing P again will enable the security code to be entered digit by digit using the Up and Down buttons to change the flashing digit and the P push button to move to the next digit If the correct code is entered pressing E will cause the first parameter ZEro to be displayed If an incorrect code is entered or if no button is pressed for ten s...

Page 19: ...s are used to make frequent calibration changes we recommend that these references should be initially checked once per year as described in sections 6 8 1 to 6 8 5 If the BA378C is always calibrated with an external reference such as a voltage source using the trim menu it will not be necessary to check the internal references 8 CALIBRATION EXAMPLE In this example a BA378C is required to display ...

Page 20: ...yed press E to store the conditioning and calibration information and return to the display mode All the display decimal points will be activated while the information is being stored This completes the calibration of the transmitter using the internal references The following steps explain how using the Trim Menu the BA378C transmitter may be calibrated against an external reference such as a vol...

Page 21: ...rn the transmitter to the SPAn prompt This completes calibration at 240 0o C Pressing the E push button again will return the transmitter to the display mode 9 MAINTENANCE The instrument should only be maintained by trained competent personnel 9 1 Fault finding during commissioning If the BA378C transmitter fails to function during commissioning the following procedure should be followed Symptom C...

Page 22: ... hold a card showing the units of measurement e g o C o F kg The transmitter can be supplied with a printed scale card showing any units specified at time of ordering If a printed scale card is not requested a blank card will be fitted Scale cards can easily be marked on site as follows a Un plug the rear terminal block s disconnect the input wires and remove the rear panel which is secured by fou...

Page 23: ... output parameters of the circuit do not exceed Uo 30V dc Io 280mA Po 0 85W The maximum equivalent capacitance and inductance between each set of alarm terminals is Ceq 40nF Leq 20µH These figures must be subtracted from the maximum permitted cable capacitance and inductance specified by the system certificate of the circuit connected to the alarm Fig 12 Typical application of alarms 11 1 3 Progra...

Page 24: ...ltering any of the alarm parameters To check or change the function select EnbL from the alarm sub menu and press P which will reveal the current setting The function may be changed by pressing the Up or Down button followed by the E button to return the EnbL prompt 11 3 2 Setpoint adjustment SP1 and SP2 The setpoint of each alarm relates to the transmitter display and each may be independently po...

Page 25: ...25 ...

Page 26: ...3600 seconds in 1 second increments To adjust the alarm silence time select SIL from the alarm sub menu and press P which will reveal the existing time Each digit may be adjusted using the Up and Down push buttons and the P button to move to the next digit When the required time has been entered press E to return the SIL prompt 11 3 8 Access Setpoint AcSP This function controls the parameters in a...

Page 27: ...he next digit When the required setpoint has been entered pressing E will return the display to the SP1 or SP2 prompt from which the other setpoint may be selected or the transmitter may be returned to the display mode by pressing E again Direct access to the alarm setpoints is only avail able when the menu is enabled see section 11 3 8 12 BACKLIGHT The BA378C can be supplied with LED backlighting...

Reviews: