7
2.3 Displays
The BA384NG has two digital displays and
associated annunciators, plus a flow indicator as
shown on the front cover of this manual.
Total display
Shows the total flow of the
selected input or the composite
flow A + b or A - b usually on the
upper eight digit display. May
be reset to zero by the front
panel push buttons or a remote
reset switch.
Rate display
Shows the flow rate of the
selected input or the composite
flow rate A + b or A - b usually on
the lower six digit display.
Flow indicator
This disc in the lower left hand
corner of the instrument display
'rotates' for two seconds each
time an input pulse is received
on the input being displayed.
Appears to rotate continuously
when input frequency exceeds
0.5Hz.
Hold
Activated when input frequency
annunciator
is below the clip-off threshold for
the input being displayed.
Reset
Activated while the instrument is
annunciator
being reset.
Rate
Identifies the rate display
annunciator
Total
Identifies the total display
Annunciator
RTx
Retransmitted pulse
annunciator
annunciator.
Depends upon the setting of
5ource
in the pulse output
configuration menu.
5caled:
Annunciator activated each
time pulse output open
collector is
on
, i.e. Ron is
less than 60
Ω
+ 3V.
direct:
Annunciator
continuously
activated.
2.3.1 Display over-range
Over-range of the upper eight digit display or the
lower six digit display is indicated by all the digits
displaying
9
and all the decimal points flashing.