background image

34

Select the required sub-function

oP2

 

:

 

on1

Time oP2 turns 

on

 first time

oP2

 

:

 

oFF1

Time oP2 turns 

off

 first time

oP2

 

:

 

on2

Time oP2 turns 

on

 second time

oP2

 

:

 

oFF2

Time oP2 turns 

off

 second time

When  selected  pressing 

(

  will  show  the  existing

time with the hours flashing.     Using the 

*

  or 

&

button adjust the hours and then press 

(

 to transfer

control to the minutes display, pressing 

(

 again will

transfer control to the seconds display.     When the
time has been set as required press 

)

 to return to

the sub-menu from which another 

on

 / 

off

 time may

be selected for adjustment.     When all the  

on

  /  

off

times have been entered, return to the configuration
menu by pressing the 

)

 button twice.

8.4.13 Access control output 

on

 & 

off

  

times

             from display mode:  

AC5P

This   function   activates   a   separate   menu   that
provides direct access to the control output's 

on 

and

off

  times   when   the   Clock   is   the   display   mode

(displaying time).   An operator may therefore adjust
the  

on

  and  

off

  times  without  having access  to the

instrument configuration menu.  Further protection is
provided   by   a  separate   security   code.     When   this
function is enabled the  

on

  and  

off  

times of the two

control   outputs   may   be   adjusted   from   the   display
mode by simultaneously operating the 

(

  and 

*

push   buttons.     The   function   contains   two   sub-
functions, 

EnbL

  which   activates   the   function   and

ACCd

 which defines a separate access code that may

be   used   to   prevent   the  

on

  and  

off

  times   being

accidentally adjusted from the display mode.

To check or change the function, select 

AC5P

  in the

configuration menu and press 

(

  which will reveal

the 

EnbL

 prompt, pressing 

(

 again will show if the

function is 

on

  or 

oFF

.    If adjustment of the control

output times from the display mode is not required
press the 

*

  or 

&

  button to select 

off

  and then

press 

)

  twice to return to the configuration menu.

If the function is required, select 

on

 and press 

)

 to

return to the 

EnbL

  prompt from which 

ACCd

, which

allows a separate access code to be entered, can be
selected by pressing the 

*

 or 

&

 button.

Access to the control output times from the display
mode may be protected by a four digit alphanumeric
security code which must be entered to gain access.
Default   security   code 

0000

  allows   unrestricted

access.   With 

ACCd

  displayed, press 

(

  to enter a

new   access   code.     The   BA577E-SS   Clock   will
display 

0000

  with   one   digit   flashing.    The   flashing

digit   may   be   adjusted   using   the 

*

  or 

&

  push

button, when set as required operating the 

(

 button

will transfer control to the next digit.     When all the
digits have been adjusted press 

)

 twice to return to

the       

AC5P

  prompt in the configuration menu. The

revised   security   code   will   be   activated   when   the
BA577E-SS is returned to the display mode.

8.4.14 Security code:   

CodE

Access to the instrument's configuration menu may
be   protected   by  a  four  digit   alphanumeric   security
code which must be entered to gain access.     New
instruments are configured with the default security
code  

0000

  which   allows   unrestricted   access   to   all

configuration functions.

To enter a new security code select  

CodE

  from the

configuration menu and press  

(

  which will cause

the Clock to display 

0000

 with one digit flashing.  The

flashing digit may be adjusted using the 

*

 and 

&

push   button,   when   set   as   required   operating   the

(

  button   will   transfer   control   to   the   next   digit.

When all the digits have been adjusted press 

)

 to

return to the 

CodE

 prompt. The revised security code

will be activated when the Clock is returned to the
display mode.  

Please contact BEKA associates sales department if
the security code is lost.

8.4.15 Reset   configuration   to   Clock   factory  

defaults:.   

r5Et dEF

When the BA577E-SS is configured as a Clock, this
function resets the configuration to the Clock factory
defaults shown in sections 8.4 of this  manual.

To   reset   the   instrument   configuration   select

r5Et dEF

 from the configuration menu and press 

(

.

The Clock will display 

0000

 with the first digit flashing

which   is   a   request   to   confirm   the   instruction   by
entering  

5urE

.   Using the  

*

  or  

&

  button set the

first   flashing   digit   to  

5

  and   press  

(

  to   transfer

control to the second digit which should be set to 

u

.

When 

5urE

 has been entered pressing the 

)

 button

will reset all the configuration functions to the factory
defaults   and   return   the   instrument   to   the   display
mode as a Clock with default configuration.

Summary of Contents for BA577E-SS

Page 1: ...Issue 6 4th May 2021 BA577E SS Rugged one input general purpose Timer or Clock Issue 6...

Page 2: ...t 5 5 11 Repeat timing cycle CYCLE5 5 5 12 Cycle function enable EnbL 5 5 13 Cycle count cycl cnt 5 5 14 Restart delay r5t dela 5 5 15 Adjusting the set time 5Et t and restart delay r5t dela from the...

Page 3: ...4 8 Enable Control output 1 EnbL 8 4 9 On and off times oP 1 on 1 oP 1 off 1 oP 1 on 2 oP 1 off 2 8 4 10 Control output 2 Optional oP2 8 4 11 Enable control output 2 EnbL 8 4 12 On and off times oP2...

Page 4: ...circuit that is used for all BA577E SS Timer installations Clock installations are the same except that the input terminals 5 6 are not used For simplicity the optional control outputs are described s...

Page 5: ...etic pick off sensor which is not suitable for controlling the BA577E SS 2 2 5 Voltage input Two voltage input ranges are independently selectable in the BA577E SS Timer configuration menu VoLt5 L and...

Page 6: ...BEKA stainless steel support plate should be slid over the rear of the Counter before the panel clamps are fitted to evenly distribute the clamping force and prevent the enclosure panel being distort...

Page 7: ...g the requested units of measurement if this information is not supplied when the instrument is ordered a blank card will be fitted A pack of self adhesive scale cards printed with common units of flo...

Page 8: ...isplay through which a scale card showing the units of time such as hours can be seen New instruments are fitted with a scale card showing the units of measurement specified when the instrument was or...

Page 9: ...ished via the front panel push buttons or by a remote contact A grand total time is maintained which can be viewed by operating the front panel push buttons It may be reset to zero from within the con...

Page 10: ...5 Shows in succession firmware version number instrument function elap5e and any output accessories that are fitted A Dual Control Outputs Access to configuration menu 5 3 Displays when configured as...

Page 11: ...d of the delay time automatically returns to Running Complete Entered automatically from Running when there are no more timing cycles to perform When the CYCLE5 function is enabled the timing cycle ca...

Page 12: ...he instrument using the or push button to adjust the flashing digit and the push button to transfer control to the next digit If the correct code has been entered pressing will cause the first paramet...

Page 13: ...itch contact Energise input by linking terminals 3 4 See section 5 5 5 dEbounCE Defines the level of debounce applied to the input to prevent false functioning dEFAuLt HEAVY LiGHt See section 5 5 6 Di...

Page 14: ...e display mode to 5Et t and r5t dela See section 5 5 15 Display Summary of function uP or dn Direction of count Defines whether the Timer times down from 5Et t to zero or times up from zero to the set...

Page 15: ...control output 1 See section 5 5 26 Display Summary of function CLr Gtot Resets grand total to zero This function resets the grand total which represents the total Timer run time from within the conf...

Page 16: ...s required press twice to return to the configuration menu or repeatedly press the or button until the required type of input is displayed and then press twice to return to the configuration menu One...

Page 17: ...n at start of each period 02 11 Current Number of cycles cycle requested not shown number when cycle is continuously repeated LabEL Exactly the same as 5td but timer status i e CYCLE or dELAY is shown...

Page 18: ...18...

Page 19: ...19...

Page 20: ...l to the next digit When set as required enter the selection and return to the cycl cnt prompt in the sub menu by operating the button 5 5 14 Restart delay r5t dela This sub function defines the time...

Page 21: ...mediately Continue When power is restored the Timer will continue from where it stopped without any manual intervention To check or change the function select P fail from the configuration menu and pr...

Page 22: ...uration example in section 6 2 of this manual illustrates the use of the control outputs Further examples can be found in Application Guide AG374 which is available on the BEKA website 5 5 22 Control...

Page 23: ...it can be adjusted using the or button and the button to move to the next digit When the required delay has been entered press to return to the control output 1 sub menu The Timer configuration exampl...

Page 24: ...rol to the second digit which should be set to u When 5urE has been entered pressing the button will reset all the configuration functions and return the instrument to the display mode Note r5Et dEF d...

Page 25: ...he required instrument configuration for this example are shown below Function Display Setting Access code CodE 0000 Function FunCtion ELAP5E Input inP type ContACt debounce dEbounCE dEFAuLt Display 2...

Page 26: ...r configuration menu Incorrect security code That the correct security code is being used Contact BEKA if code is lost 7 2 Fault finding after commissioning ENSURE PLANT SAFETY BEFORE STARTING MAINTEN...

Page 27: ...e regularly checked Inspection frequency should be chosen to suit the environmental conditions 7 5 Guarantee Instruments which fail within the guarantee period should be returned to BEKA associates or...

Page 28: ...rument s permanent memory Following initialisation the instrument will display a flashing display which is a request for the local time to be entered The clock will not start to function until a displ...

Page 29: ...hroughout this manual push buttons are shown as and and legends displayed by the Clock are shown in a seven segment font as they appear on the instrument e g di5PLAY and CodE Access to the configurati...

Page 30: ...30...

Page 31: ...ntrol output 1 Optional Control output 1 can turn on output closed and off output open twice in each twenty four hour period This function contains five sub function EnbL Enables output 1 and defines...

Page 32: ...et time is entered the Clock will display a flashing 00 00 00 or 00 00 for 24 hour formats or A12 00 00 or A12 00 for 12 hour formats To adjust the clock set time select 5Et from the configuration men...

Page 33: ...tered return to the configuration menu by pressing the button twice 8 4 10 Control output 2 Optional oP2 Control output 2 is an optional factory fitted output which can be configured to turn on output...

Page 34: ...e display mode may be protected by a four digit alphanumeric security code which must be entered to gain access Default security code 0000 allows unrestricted access With ACCd displayed press to enter...

Page 35: ...g will then change the instrument to a Clock and return the display to FunCtion in the configuration menu Step 3 Select display format Using the or button select di5PLAY in the configuration menu and...

Page 36: ...access to the configuration menu Using the and buttons select CodE from the configuration menu and press which will reveal 0000 with the first digit flashing This example requires the security code to...

Page 37: ...ANCE If a BA577E SS fails after it has been functioning correctly as a Clock the following table may help to identify the cause of the failure Symptom Cause Check No display No power supply That there...

Page 38: ...8 10 6 Customer comments BEKA is always pleased to receive comments from customers about our products and services All communications are acknowledged and whenever possible suggestions are implemented...

Reviews: