Installation
METPOINT
®
OCV
21
Basic requirements regarding the installation:
The measuring gases must be free from dust and water
when supplied to the METPOINT
®
OCV measuring device.
Therefore, gas processing adjusted to the measurement task
is required in most application cases.
Choose a clean, dry and dust-free room which is protected
against storm.
•
The METPOINT
®
OCV must not be employed in a
hazardous area!
•
The measuring section must meet the following
requirements:
- centric probe seating
- made of stainless steel
- oil- and grease-free
- dimensions in accordance with the BEKO
provisions
- inflow towards the probe from the bottom to the top
(see flow direction)
•
The measuring section must be installed vertically.
•
The employment of an original BEKO measuring
section is recommended.
•
The installation of a bypass (see Chapter 10.5) or of
a ball valve between the sensor unit and the
measuring section is recommended.
•
Ambient temperature (sensor unit) +5°C to 45°C.
•
Ambient temperature (evaluation unit with user
interface)
+5°C to +50ºC.
•
The maximum pressure dew point of +10°C must
not be exceeded.
•
Other connection possibilities and measuring
methods are possible, depending on the respective
application. Please contact us in this regard.
Measuring
gas inlet