RTD THERMOMETER
1
1. INSTRUCTION
This instrument is a digital thermometer for use with platinum-type temperature sensor.
Temperature indication follows IEC751 temperature table for PT-type sensor.
2. SPECIFICATIONS
Numerical Display:
5 digital liquid crystal displays
Measurement Range:
-100°C to 400°C; -148°F to 752°F
Resolution:
0.01°C; 0.02°F;
Sensor types:
Platinum resistance temperature sensor for pt-100 4 wires.
ALPHA=0.00385
Environmental:
Operating Temperature and Humidity:
0°C to 50°C (32°F to 122°F) ; 0 to 80% RH
Storage Temperature:
-10°C to 60°C (14°F to 140°F); 0 to 80% RH
Altitude up to 2000 meters.
Accuracy: at ( 23 ± 5°C; 73± 9°F )
Range
Accuracy
-100°C ~ 400°C
±(0.05% r 0.1°C)
-148°F ~ 752°F
±(0.05% r 0.2°F)
Temperature Coefficient:
For ambient temperatures from 0°C ~ 18°C and 28°C ~ 50°C, for each °C ambient below 18°C or
above 28°C add the following tolerance into the accuracy spec.
0.005% of r 0.01°C ( 0.005% of r 0.02°F )
Note:
The basic accuracy Specification does not include the error of the probe please refer to the
probe accuracy specification for additional details.
Sample Rate:
2 times per second
Dimension:
185 x 65 x 36 mm (meter only)
Weight:
360g Approx.
Power requirement:
9V Battery;
AC adapter: 9V DC(7 to 10V Max.)/20mA Min (Plug Diameter: 3.5mmx1.35mm)
Battery Life:
Approx. 100hours
Standard Accessory:
Pt-100 Probe (class A)
,
Instruction manual, 9V Battery, carrying case.
Optional Accessories:
AC adapter.
Summary of Contents for B60901-0100