RTD THERMOMETER
7
8. TEMPERATURE METER CALIBRATION SETUP
The below table diagram shows each button function when in the
calibration mode.
Note: Turn the Auto Power-Off off before attempting the following
SETUP. Setup mode is cancelled if the
button is pressed during
the procedure.
8.1 Temperature Calibration
Place the temperature probe in a known, stable temperature environment. Allow the reading to
stabilize. In the “Setup” mode, change the offset until the display reading matches the calibration
temperature.
8.1.1
To enter the setup mode, press and hold the
and
button, then power on the meter
8.1.2
Press the
button to enter the calibration mode. (in 3 seconds)
8.1.3
Press the
button to enter the calibration mode. (in 3 seconds)
8.1.4
Press the
or
button to confirm the present Temperature value.
8.1.5
Press the
button twice to confirm selection.
Note: To abort during the process, press the
button to exit calibration mode.
8.2 Recall Default factory setting value
8.2.1
To enter the setup mode, press and hold the
and
buttons, then power on the meter.