PX106545-EN, Rev.A
59
7.5
Loss limit over
Error message generated when the estimated loss value is higher than the preset loss
factor limit.
Solution: Check the loss factor limit.
7.6
Fiber thin error
Error message generated when the spliced point becomes thinner than the standard after
splicing.
Solution: Make an adjustment to shorten the pulling length of the pulling splice. Check
whether the arc amount or arc time is set as too large or too long.
7.7
Fiber thick error
Error message generated when the spliced point becomes thicker than the standard after
splicing.
Solution: Reduce the overlap set value. Check whether the arc amount or arc time is set
as too small or too short.
7.8
Bubble error
Error message generated when there are bubbles or spots being generated on spliced
point after splicing.
Solution: Examine the fiber cleaver. Clean the V-groove. Examine the electrodes.
7.9
Cleaved surface error
Error message generated when the cut surface of the fiber is of poor
quality. Solution: Check the condition of the fiber cleaver. Re-cleave
the fiber.