10.3
6000
V
ISIBILITY
S
ENSOR
R
EMOVAL
P
ROCEDURES
:
SAFETY PRECAUTIONS
2. Do not perform outdoor Belfort AWOS maintenance tasks during a thunderstorm.
3. Use extreme caution when lowering the tilt-over tower whenever ice or snow has accumulated
on the tower cross arm or the sensors. The extra weight may cause the tower to fall over
rapidly.
TOOLS REQUIRED
1.
8 foot step ladder.
2.
Socket wrench with 9/16" socket.
.
Removal Procedures: Refer to Figure 7.
1.
At the DATU display terminal, issue a NOTAM advisory to alert users that the Belfort
AWOS system is being serviced prior to beginning investigation of this fault. The
NOTAM to be issued is "
Belfort AWOS system being serviced, do not use current
weather data outputs
." Verify that this NOTAM advisory is displayed on the operator
screen prior to beginning this maintenance tasks. Refer to the "Belfort AWOS Operator
Manual" for instructions on how to enter and delete NOTAMs.
2.
Remove tower-securing bolts and tilt the tower until the cross arm is at a comfortable
working height to access the Temperature/Humidity Sensor. Secure the tower in that
position by tying the attached tower hold-down rope to the base of the Belfort AWOS tower
as shown in Figure 2.
3.
Power down the power enclosure by opening the AC and DC circuit breakers.
4.
Inside of the power enclosure, disconnect the red and black visibility power harness
connections from the terminal block TB3.
5.
Inside of the power enclosure, disconnect the visibility sensor RS232 serial cable that are
connected to the terminal block TB4
.
6.
Remove the screws that hold the conduit clamps which secure the sensor wire bundle
running down the hinge side of the tower. Save the screws and clamps.
7.
Remove the mounting bolts that secure the visibility sensor and Ambient Light Sensor to
the tower and carefully remove the sensors from the tower.
8.
Inspect the sensor cables and connectors for damage, corrosion or bent or broken pins.
Replace damaged cables as necessary.
9.
Contact Belfort Instrument Co. customer support to make arrangements for replacement of
defective components.
10.
Refer to the Belfort AWOS Installation and Checkout manual Section 9.0 for installation
and checkout of the replacement parts.
11.
Document sensor replacement on the FAA form 6030-1.
Summary of Contents for AWOS
Page 17: ... 14 BELFORT INSTRUMENT Proprietary Information Figure 2a Securing the Tower for Maintenance ...
Page 49: ... 46 BELFORT INSTRUMENT Proprietary Information Figure 9 Temperature Humidity Sensor Removal ...
Page 51: ... 48 BELFORT INSTRUMENT Proprietary Information Figure 10 Visibility Sensor Removal ...
Page 60: ...BELFORT INSTRUMENT Proprietary Information Appendix A Belfort AWOS Maintenance Forms ...
Page 66: ...BELFORT INSTRUMENT Proprietary Information This page intentionally left blank ...
Page 74: ...BELFORT INSTRUMENT Proprietary Information Appendix C UHF Transceiver Specifications ...
Page 88: ...BELFORT INSTRUMENT Proprietary Information Appendix F DCP Site Configuration Procedure ...