11 System Recovery
S008459 Rev C
Page
113
of 117
Because of the elastic rubber elements in the compensator, it may be necessary to fasten
them temporarily.
Figure 11.2: Compensators lifted
5. Insert the Recovery Key into the serial port on the control panel
Figure 11.3: Inserting the Recover Key
6. Power ON the scanner
The scanner will find the Recovery Key present in the serial port and will automatically
execute the recovery sequence. All viable system data will be restored and necessary
repair steps will be taken without the need of any user interaction.
Caution
The recovery sequence can last up to two minutes. While the recovery sequence is
running,
no message
will be displayed. DO NOT power OFF the scanner at any time
during the recover procedure!
7. When the recovery sequence has finished, the scanner will power OFF automatically
8. Unplug the Recovery Key after the scanner has powered OFF
9. Set the compensators into their original position and reinsert the pressure plate
10. Power ON the scanner and using a host PC connected to the scanner via the supplied
crossover cable, launch the Infinity Web-based Scan Application using an Internet browser