Adjustment after exchanging the power PC board (mA adjustment)
1. Turn on the main switch at the circuit breaker panel and the main switch on the x-ray front panel.
2. Hold down the tooth selection switches together (T1, T4, and T5) until “h.xx” appears in the exposure time
display window.
3. Wait until the display changes to “0.01.” Make an x-ray exposure by depressing the exposure switch.
4. Repeat step 3 until “Fin” is displayed in the exposure time display window (you will make approximately
eight (8) to ten (10) x-ray exposures until “Fin” is displayed.
5.
Turn off the main power switch on the front panel of PHOT-X II.
-28-