Doc 0256680 rev L
Page 21 of 22
Not subject to issue control when printed
HI Testing
Open batt door and insert a batt pill. Close batt door and attach the prog adapt and the prog cable to the HI. It is recommended to use the prog clip
during HI test to prevent the prog cable to fall off - please see the “Prog adapt clip” page
Mount the 2cc coupler to the TM12 coupler mic. Insert the test rec into the white fixture. Attach the white fixture to the black test frame and place it all
in the test chamber as shown. Connect the HI to the rec. Select the correct test prog and make sure to follow instructions coming up on the screen
ATC4000 sound chamber prepared
for test of HI
Note about PROG. CABLE EXTENSION p/n 17517800:
The extension cable is needed for making correct current measurement during DSA6000 HI test
on some platform models. Testing these devices without the extension cable will show an incorrect
current measurement. However, if the ext cable is used during a DSA6000 HI test of a model where
the ext cable is not needed, it will not affect the test. Therefore, to avoid failures and misunderstand
-
ings it is safer just to leave the ext cable on the DSA at all times and use it during test of all devices,
no matter if it is required or not
Connect the extension cable to the
prog cable and the DSA as shown