Delayed Activation
Your hearing care professional may have activated the Delayed Activation function in your hearing instru-
ment. The Delayed Activation function delays the turn-on time by 10 seconds after closing the battery door.
This is indicated by beeps in the receiver at one second intervals when the battery door is closed, meaning
that 10 beeps will be heard before the hearing instrument is activated.
If you wish, this function can be deactivated by your hearing care professional.
Battery door lock
Your hearing care professional can install a Left/Right side indicator (Blue=Left, Red=Right), which can be
used to lock the battery door. Once locked, the instrument can only be turned off once the instrument is
unlocked.
To lock the battery door:
1. Close the battery door
2. Using the battery door locking tool, slide the color marker to the right.
To unlock the battery door:
Slide the color marker to the left.
Unlock
Lock
Low Battery Indication and changing batteries
Low battery indication
Your hearing care professional can set your hearing instrument to give
an acoustical indication when the battery is reaching its end of life. The
hearing instrument will reduce amplification and emit a melody if battery
power gets too low. This signal will recur every 5 minutes for OR65/75/85
models, and every 15 minutes for BO95 model, until the hearing instru-
ment automatically switches off. It is recommended that you keep spare
batteries on hand.
Changing batteries
Open the battery compartment completely by using your fingernail. Use
the magnetic end of the cleaning brush for easy battery removal and in-
sertion.
When replacing the battery, always use new Zinc-Air batteries that have a
minimum remaining shelf-life of 1 year.
The 65 model uses a size 312 battery.
The 75 and 85 models use a size 13 battery.
The 95 model uses a size 675 battery.
open battery door completely
exchange battery