Important information
40
isoHV425_D00082_03_M_XXEN/05.2020
7.5 Device-specific data types
7.5.1
Device name
The data format of the device name is specified below.
7.5.2
Measured values
Each measured value is available as a channel and consists of 8 bytes (4 registers). The
first measured value register address is 1000. The structure of a channel is always iden-
tical. Content and number depend on the device. The structure of a channel is shown
with the example of channel 1:
7.5.2.1
Float = Floating point value of the channels
Representation of the bit order for processing analogue measured values according to
IEEE 754
S = Sign
E = Exponent
M = Mantissa
Word
0x00
0x01
0x02
0x03
–––––––––––––––––––
0x08
0x09
10 words in total
Each word contains two ASCII characters
1000
1001
1002
1003
HiByte LoByte HiByte LoByte HiByte
LoByte
HiByte
LoByte
Floating point value (Float)
Alarm type
and test
type (AT&T)
Range and
unit (R&U)
Channel description
Wo
rd
0x00
0x01
Byt
e
HiByte
LoByte
HiByte
LoByte
Bit
31
30
24
23
22
16
15
8
7
0
S E E E E E E E E M M M M M M M M M M M M M M M M M M M M M M M