21
Note :
●
Please read this user’s manual and instructions for control solution test
carefully before use.
●
Do not reuse the test strips and always close the cap of test strip vial tightly
after used.
●
The first three drops of control solution must be discarded and the vial should
be shaken well before use when the vial is newly opened.
●
Newly opened vials of control solution and test strips must be marked with an
opening date.
●
Do not use BeneCheck PLUS Multi-Monitoring System to do any test until you
can get the control test result to fall within the control range listed on the test
strip vial.
●
The test results fall outside the control range, please reconfirm and repeat the
test, if it still fall outside the range, please contact an authorized distributor
for help and service.
●
Uric acid control solutions and total cholesterol control solutions are not
included. Please contact an authorized distributor for ordering.