Chapter 8 Error Message and Trouble Shooting
Message
Cause
Solution
E-0
There will be a problem
with the test strip.
Repeat the test with a new test
strip. If the problem persists, please
contact the dealer for service.
E-b
The power of the battery
is too low to run a test.
Replace the battery at once.
Err.3
The reference voltage or
other circuit is error.
Call local authorized distributor
for help and service.
E-E
1. There may be a problem
with self-check of the
meter.
2. Use of the wrong or
damaged code strip
or the code strip was
inserted improperly.
Power off then retry, if the error
appears again, contact the authorized
distributor for service.
Check that the code number on the
meter display matches the code
number on the test strip vial. Code
the meter or retest as necessary. If
the error message appears again,
please contact the dealer for service.