03/ 2011
BENNING MM 11
29
- Remove the rubber protection frame
from the BENNING MM 11.
- Place the BENNING MM 11 onto its front and turn the slotted screws of the
battery compartment by 90 ° in an anticlockwise direction.
- Lift the battery compartment cover off the bottom section.
- Lift the battery from the battery compartment and carefully disconnect the
battery supply lines from the battery.
Do not undo any screws on the printed circuit. Leave the
printed circuit in the bottom section of the housing. Danger for
the interface LED!
- Remove the top screw (black) and the two screws in the battery compart-
ment from the bottom section (housing bottom).
- Now place the unit on its back again.
- Lift the front end at its bottom section and carefully remove the same.
- Lift one end of the defective fuse from the fuse holder.
- Fully push the defective fuse out of the fuse holder.
- Fit the new fuse with the same rated current, rated voltage, same discon-
nection capacity, same trigger characteristics, and same dimensions.
- Make sure that the new fuse is seated centred in the holder.
- Order the battery supply lines such that they are not crushed between the
housing parts.
- Engage the front end on the housing bottom and fit the three screws.
- The battery must be connected to the battery supply lines and placed in the
position provided for this purpose in the battery compartment.
- Engage the battery cover into the bottom section and turn the slotted
screws of the battery cover by 90 ° in a clockwise direction.
- Place the BENNING MM 11 into the rubber protection frame
See figure 13:
Fuse replacement
9.5 Calibration
To maintain the specified accuracy of the measurement results, the instrument
must be recalibrated at regular intervals by our factory service. We recommend
a recalibration interval of one year. Send the appliance to the following address:
Benning Elektrotechnik & Elektronik GmbH & CO. KG
Service Centre
Robert-Bosch-Str. 20
D - 46397 Bocholt
9.6 Spare parts
Fuse F 1 A, 600 V, 10 kA, D = 10 mm, L = 35 mm,
part no. 749598
Fuse F 15 A, 600 V, 100 kA, D = 10 mm, L = 38 mm,
part no. 749595
10. Application of rubber protection frame
- You can store the safety measuring leads by winding the safety measuring
leads around the rubber protection frame
and engaging the tips of the
safety measuring leads with protection against the rubber protection frame
(see Figure 14).
- You can engage a safety measuring lead to the rubber protection frame
such that the measuring tip protrudes freely, in order to be able to apply the
measuring tip - jointly with the BENNING MM 11 - to a measuring point.
- The rear support on the rubber protection frame
provides for an inclined
set-up of the BENNING MM 11 (facilitates reading) or attachment (see
Figure 15).
- The rubber protection frame
is fitted with a lug which can be freely used
for an attachment option.
See figure 14:
Winding up of the safety measurement lead
See figure 15:
Erecting the BENNING MM 11
11. Technical data of the measuring accessories
- Standard: EN 61010-031,
- Maximum rated voltage to earth (
) and measuring category:
With push-on caps: 1000 V CAT III, 600 V CAT IV,
Without push-on caps: 1000 V CAT II,
- Maximum rated current: 10 A,
- Protective class II (
), continuous double or reinforced insulation,
- Contamination class: 2,
- Length: 1.4 m, AWG 18,
-
Environmental conditions:
Maximum barometric elevation for making measurements: 2000 m,
Temperatures: 0 °C to + 50 °C, humidity 50 % to 80 %
- Only use the measuring leads if in perfect and clean condition as well as
according to this manual, since the protection provided could otherwise be