BV200 Plasma Viscometer Mk2/3 Customer User
Manual
Issue 9
th
Sep 2020 - Rev 1.8
Page 39 of 53
6.4. Manual decontamination process
The manual decontamination cleaning process comprises 7 stages.
Warning:
Do not routinely use bleach products when cleaning/decontaminating the viscometer
as this can impact the serviceability of the viscometer system components.
We recommend using Benson Viscometers barcoded decontamination additive, shown below.
From the Maintenance Menu on the main screen select do Decontamination Clean
1.
The following popup will be displayed, with the following text:-
2.
Add the barcoded decontamination additive to 118ml of demineralised.
Screw on the Decontamination bottle cap.
Remove the wash tube from the top of the Wash bottle and connect it into the tube
adaptor on the Decontamination bottle cap.
a.
Replace and screw on the decontamination bottle cap.
b.
Remove the wash tube from the top of the wash bottle and connect it into the tube
adaptor on the decontamination bottle cap.