will remove the recently created sensor unit from the drop down menu.
Sensitivity mV/SU:
This is a user selectable input value. Enter the sensitivity
of the device you are connecting to in the input box. Use the [
Navigation
Keypad
] and enter the value via the keypad and press [
OK
] when complete.
Input Range
: Select Auto, Up Only, 10 mV, 20 mV, 20 mV, 50 mV, 100 mV, 200
mV, 500 mV, 1V, 2V, 5V, 10V, or 20V. This should match the output of the device
that you are connecting to. Up only is typically used with modal Fieldpaq testing
where the range will increase “up” as required to locate the correct input range.
Auto will automatically collapse the input range around the center. Use this if you
are unsure of the correct input range.
Coupling
: Select AC, DC, IEPE, 200V Mic. or
0V Mic.
Soft Key Menu: Display
The display function will determine what is
displayed. You can choose between Time
Waveform and Spectrum. Select 100, 200,