CLOSED ORBIT OPERATING MODE
This is the most simple operating mode. It is the default mode, i.e. when no external control
signals are applied to the BPM module.
In the default mode, the internal clock controls the button sampling frequency (8…10 kHz).
The average beam position is available as:
XOUT
X output. Analog output. (See Algorithm & Polarity Conventions, this manual)
Voltage range: –10V…+10V. Zero is centered beam.
YOUT
Y output. Analog output. (See Algorithm & Polarity Conventions, this manual)
Voltage range: –10V…+10V. Zero is centered beam.
The value of each individual button A, B, C and D is also available. They can be used to compute
–externally– the beam position with algorithms other than
∆
/
Σ
:
AOUT
A output. Analog output.
Voltage range: 0…+10V. Pedestal ca. 1 V.
BOUT
A output. Analog output.
Voltage range: 0…+10V. Pedestal ca. 1 V.
COUT
A output. Analog output.
Voltage range: 0…+10V. Pedestal ca. 1 V.
DOUT
A output. Analog output.
Voltage range: 0…+10V. Pedestal ca. 1 V.
In closed orbit mode, the BPM module performance can be monitored with:
PLLOCK
PLL demodulator in lock & good signal level. TTL output.
High = PLL in lock and signal level OK
Low = PLL out of lock or signal level poor
The DB9 female front-panel connector allows the observation of the multiplexed demodulated
signal:
SDEMOD
Synchronous demodulator output. Analog output. For oscilloscope viewing.
SYNC
Synchronization pulse. TTL output. Positive-going. For oscilloscope trigger.
BERGOZ Instrumentation
Multiplexed
01630 Saint Genis Pouilly, France
Beam Position Monitor
Tel. +33-450.426.642
Fax +33-450.426.643
Page 13
User’s manual
BERGOZ Instrumentation - 01630 Saint Genis Pouilly, France - Tel.: +33-450.426.642 - Fax: +33-450.426.643
TVA-VAT-IVA-USt. Nº FR88414997130 - Sàrl. capital 152K - Siren 414 997 130 R.C.S. Bourg - APE 332B