X AND Y GAIN ADJUSTMENT
Conventions
BPM modules are referenced in terms of operating frequency and X / Y gains as they appear on the
order code:
BPM- . MHz-X . V/%-Y . V/%
Operating frequency
Horizontal gain
Vertical gain
The BPM module X / Y gain units are V/%.
The BPM buttons sensitivity is -depending on the laboratory- expressed as:
• Vacuum chamber calibration factor [mm]
• Vacuum chamber Sensitivity [mm
-1
] or [%/mm] or [dB/mm]
The overall gain of the BPM buttons and electronics system is expressed in V/mm.
It is the product:
Gain [V/mm] = Vacuum chamber Sensitivity [%/mm] x BPM module gain [V/%].
BPM module gain adjustment
The X and Y gains have been adjusted before shipment according to the order code.
Two on-board potentiometers allow a fine adjustment of the gain: X gain and Y gain:
BUTA
BUTB
BUTC
BUTD
IF BPF
Level shift
AGC
IF BPF
RF BPF
Attenuators
Clock
adjust
Y gain
Σ
adj.
X gain
LO
Lock
Carrier phase
F-Key
Demod. phase
Demodulator
SBS-mode
fixed level gain
GaAs
switch
driver
GATE
(option)
Single
Button
Sampling
(option)
Mixer
LNA
IF
IC
The gain potentiometers are accessible via holes bored through the Single Button Sampling (option)
piggy-back board.
BERGOZ Instrumentation
Multiplexed
01630 Saint Genis Pouilly, France
Beam Position Monitor
Tel. +33-450.426.642
Fax +33-450.426.643
Page 18
User’s manual
BERGOZ Instrumentation - 01630 Saint Genis Pouilly, France - Tel.: +33-450.426.642 - Fax: +33-450.426.643
TVA-VAT-IVA-USt. Nº FR88414997130 - Sàrl. capital 152K - Siren 414 997 130 R.C.S. Bourg - APE 332B