2) Activation of the Q-tag
®
Quad
Please always verify the expiry date prior to using any device (see "Expiry
Code Explanation").
The device is activated by pressing the START-
button for at least 2 seconds.
The active recording is shown by the
(OK symbol)
and the flashing dot on the display.
Note (only for standard Version):
Prior to start make sure that the device is not yet
activated, otherwise it will stop.
To avoid accidental deactivation of the device, the
stop function is locked for about 1 minute after the
activation.
3) Start delay time (optional)
If a start delay time has been programmed, it will be activated the moment
you press the START-button for at least 2 seconds.
During
the start delay
time
the tick, the dot (running indicator) and possibly the alarm flag will
flash alternatively.
The start delay time is not visible.
Temperature
recording
will only start after completion of the start delay time.
Note:
To avoid accidental deactivation of the device, the stop function is
locked during the start delay time.
4) Temperature recording
Placing the Q-tag
®
Quad
The activated monitor must be placed immediately in its predetermined
location. It is recommended to place the device as close to the supervised
goods as possible. This is an important prerequisite to ensure the
temperatures observed by the device are as close to the product
temperature as possible.
Note:
It is recommended to "precondition" the device for about 30 minutes
to the desired mean temperature before use, especially if no start delay
time has been programmed. This is to insure against false readings during
start up.