CN_ILLU_NRMeasurementT ool_BW_HI

1  Place the measurement tool over the ear, 

keeping the top line horizontal. With your eyes 
at the same level as the ear, determine which 
mark is closest to an imaginary line running 
over the top of the ear canal. In this example, 
the correct choice is size 2.

  Power Molds will have a better fi t if you 

choose a slightly longer wire length than you 

would when fi tting a Power Dome. If the top 

of the ear canal falls between two numbers on 
the gauge, choose the higher number.

2  If the speaker unit wire is too long the hearing 

instrument will fall back over the ear and the 
 microphones will point vertically instead of 
 horizontally. It is also unlikely that the receiver 

wire will fi t the contours of the ear correctly.

3  If the speaker unit wire is too short, it will 

have a tendency to move out of the ear canal, 
and the hearing instrument will lose stability 
on the ear.

Selecting the Speaker unit Length

CN_ILLU_NR_Wearing instrument2_BW_HI_r

1

2

3

Four receiver lengths are available for both the M-Speaker and the P-Speaker. No. 1 is the shortest 

while no. 4 is the longest. Before fi tting Nano RITE hearing instruments, it is necessary to determine 

the correct speaker unit length. The measurement procedure is the same for both the M-Speaker and 
the P-Speaker. 

Wire too long

Wire too short

134056_NR_Fitting_Guide_generic_US_Private_Label.indd   8

14.03.13   13:37

NaNo RITE HEaRINg SySTEmS

9

CN_ILLU_NR_SpeakerUnitAttaching4_BW_HI

attaching the Speaker unit to a Nano RITE Hearing System

Correct speaker unit attachment is imperative. 

1  Slide the connector plug into the connector 

socket of the instrument. Take care to align the 
pins in the connector plug with the socket.

2  When the connector plug is correctly aligned, 

push it gently into place.

3  Take the lock pin out of the packaging in which 

the speaker unit is delivered (tweezers make 
this easier). Insert the lock pin with the 
 tweezers.

4  Ensure that the lock pin is fully inserted into 

the hearing instrument. Use the back side of 
the tweezers tool if necessary.

5  To test the attachment of the speaker unit to 

the hearing instrument, gently pull the speaker 
wire near the connection to the hearing 
 instrument as shown. If the speaker comes 
off, then it was not attached properly.

CN_ILLU_NRSpeakerUnitAttaching2_BW_HI

CN_ILLU_NRSpeakerUnitAttaching3_BW_HI

1 – 2

4

5

CN_ILLU_NR_SpeakerUnitAttaching5_BW_HI

3

Speaker units can easily sustain internal 
damage if dropped. Before you attach the 
speaker unit, please place a dome on the 
speaker to protect it.

134056_NR_Fitting_Guide_generic_US_Private_Label.indd   9

14.03.13   13:37

Summary of Contents for Nano RITE

Page 1: ...Nano RITE Hearing Systems fITTINg guIdE Nano RITE Hearing Systems fITTINg guIdE ...

Page 2: ...cro Mold to the M Speaker Unit 13 Detaching the Micro Mold from the M Speaker 14 Detaching the Lite Tip from the M Speaker 14 P Speaker Unit for Power Fittings 15 Using the Measurement Gauge 15 Custom Fit with Power Mold 16 Attaching and Detaching Power Mold to and from P Speaker Unit 16 Changing from a Power Dome to a Power Mold 17 Changing the NoWax Filter and the WaxStop 18 WaxStop System 18 No...

Page 3: ...delivers tangible end user benefits This guide is designed to help you fit and manage Nano RITE hearing systems 1 Push button 2 Microphone 2a Second microphone 3 Connector plug 4 Wire 5 Left right indicator 6 Speaker 7 Speaker unit size marking 8 Battery drawer 9 Ear grip 10 NoWax filter Nano RITE Hearing Systems overview 8 CN_ILLU_NRInstrumentwithSpeaker_BW_HI 1 2a 2 3 4 10 5 7 6 9 ...

Page 4: ...oldtoPowerDomeChanging1_BW_HI NaNo RITE HEaRINg SySTEmS 7 fitting Ranges Nano RITE hearing systems offer two speaker units the M Speaker which is appropriate for mild to moderately severe losses while the P Speaker provides additional power for severe hearing losses Expected fitting ranges with different ear pieces are shown below 250 500 1 000 2 000 4 000 10 000 0 20 40 60 80 100 120 250 500 1 00...

Page 5: ...E hearing instruments it is necessary to determine the correct speaker unit length The measurement procedure is the same for both the M Speaker and the P Speaker Wire too long Wire too short NaNo RITE HEaRINg SySTEmS 9 CN_ILLU_NR_SpeakerUnitAttaching4_BW_HI attaching the Speaker unit to a Nano RITE Hearing System Correct speaker unit attachment is imperative 1 Slide the connector plug into the con...

Page 6: ...1 m Speaker unit The M Speaker unit is suitable for mild to moderately severe losses and comes in four lengths It can be used for both instant and custom fittings 1 To obtain a more secure hold in the ear canal an ear grip should be used The same version works for both the left and the right ear Insert the ear grip in the slot in the receiver as shown noting the orientation If the ear grip seems t...

Page 7: ...d For dismounting a speaker from a Lite Tip NaNo RITE HEaRINg SySTEmS 13 Attaching the Lite Tip and the Micro Mold to the M Speaker Unit CN_ILLU_NRCostumMoldPspeakerAttaching3_CMYK_HI 1 Use the cup pin of the Lite Tip tool for curved receivers 2 Place the receiver into the tool and push the wire into the groove to obtain a stable attachment Make sure that the NoWax is on the receiver see page 18 3...

Page 8: ...erformance receiver which requires ample room in the ear canal To determine if your client s ear canal can accommodate this receiver please use the measurement gauge supplied This step is necessary for both instant fittings with the Power Dome and custom fittings with molds using the measurement gauge 1 Hold the handle of the measurement gauge so that the flat side of the handle is up 2 Insert the...

Page 9: ...om an impression similar to one needed for an in the canal hearing instrument It is possible to include a vent up to a maximum of 2 4 mm if specified at the time of order To remove the P Speaker unit first remove the cover by very gently sliding the special screw driver between the tube and the cover and then apply a slight upwards pressure Using the special screwdriver turn the locking plate coun...

Page 10: ... If it is clogged or there is no sound coming from the device replace the NoWax filter WaxStop System The WaxStop system is applicable for the following acoustic setups M Speaker coupled with a Micro Mold P Speaker coupled with any ear piece instant and custom solutions The WaxStop should be changed as needed to ensure efficient wax protection and to prolong the life of the speaker unit CN_ILLU_NR...

Page 11: ...INg SySTEmS 21 1 3 P SPEaKER uNIT aSSEmBLIES Part Number Designation Comments 118667 Speaker assembly P 1L For custom fit 118668 Speaker assembly P 2L 118669 Speaker assembly P 3L 118670 Speaker assembly P 4L 118662 Speaker assembly P 1R 118663 Speaker assembly P 2R 118664 Speaker assembly P 3R 118665 Speaker assembly P 4R 2 domES 2 1 oPEN domES Part Number Designation Comments 120045 Open Dome RI...

Page 12: ...22 320 00 Measurement gauge for P Speaker 118890 Lite Tip tool To be used for M Speaker with Lite Tip 4 3 2 1 Nano RITE CN_ILLU_NRCostumMoldPspeakerDetaching_BW_HI NaNo RITE HEaRINg SySTEmS 23 6 SPEaKER PaRTS Part Number Designation Comments 654 01 570 00 Dome adapter transparent For instant fit with P Speaker 506 24 030 00 Locking plate With P Speaker 543 11 770 00 Cover beige With P Speaker 543 ...

Page 13: ...03 13 BAG 134056 US subject to change ...

Reviews: