7. Startup of the HPLC System
FlowStar2 LB 514
62
7.6 Waste valve installation
Electrical connection
The FlowStar² uses the (open collector)
output labeled “Waste”
to switch the waste valve from ‘cold’ to ‘hot’ Position.
The two waste valve wires
will be connected to the “Waste” pin
and the “+12V” pin with the supply voltage.
Tubing connection
The valve has 3 tubing connectors. One is the inlet and the 2
others are the outlets. The inlet is not color labeled.
The outlet with the black color label is the so called
“cold outlet”.
This outlet is active if the status is on non radioactive waste.
It must be connected to the waste container collecting the non
radioactive waste using the tube with the black fitting.
The outlet with the red color label is the hot outlet. This one is
active if the counting level is above the defined threshold. It must
be connected to the waste container collecting the radioactive
waste using the tube with the red fitting.
Do not connect the clear tube from the inlet directly to the
FlowStar² outlet since this will dramatically increase the
background due to incoming light.
Only use tube unions with a large bore (0.75mm).
Inlet
Cold-
Outlet
Hot-
Outlet