9. Parameter Entries
FlowStar2 LB 514
78
Parameter page Configuration page 2/3
CONFIGURATION
2 / 3
LS-Pump installed:
Fraction Collector installed:
Waste Valve installed:
Printer installed:
Time:
14:02:23
Date:
17.Aug. 2007
Passwords...
Liquid Handling
Printer
System Clock
Miscellaneous
Figure 24: Configuration, parameter page 2/3
Liquid handling
LS-pump installed
Tick the checkbox (x) if you have are working with an LS pump. The re-
spective queries will appear in the methods parameter dialog only if this
checkbox is ticked. Only possible with admixture cells.
LS Pump Settings…
Select the correct pump to setup the pump communication.
“Analog” will
use the analog output signal at the “Scint Pump connector to control the
flow rate. All other settings will control the pump through the serial port at
The CPU. LB5035 is for older pumps, the LB5035-3M is the microbore
version (reduced flow rate). LB5036/5037 represents the current pump
version.
Vol. Cal. Factor
This Factor can be used to calibrate the pump to a specific scintillator.
Due to different viscosity of liquid scintillators the real flow rate can differ
from the preset flow. Using the calibration facto the flowrate can be cor-
rected to get correct flow rates. Default is 1.000
Count if pump is off
If this box is ticked, the FlowStar² will display no counts if the scintillator
pump is not running.
Data Output Delay (s)
When the previous box is ticked an additional data delay can be set when
the pump starts working. This will avoid the display of ghost peaks due
to improper mixing right after starting the pump. Useful values are be-
tween 10 and 30s, defaiult is 0. If set to 10s the FlowStar² will start dis-
playing data 10s after the pump started.
Fraction collector in-
stalled
Tick the checkbox (x) if you are working with a fraction collector.
Waste valve installed
Tick the checkbox (x) if you have a waste valve installed.
Thresholds
Luminescence warning
Defines the level fort he luminescence warning. The higher the value, the
more sensitive the luminescence warning. The luminescence warning is
disabled when entering „0“.
Printer
Printer installed
This feature is not implemented.