1. Introduction
FlowStar2 LB 514
8
The instrument is designed to be operated within a temperature range of 15 to 30
°C.
The instrument is designed to be operated at a maximum relative humidity of 80 %
for temperatures up to 31 °C decreasing linearly to 50 % relative humidity up to 40
°C.
BERTHOLD TECHNOLOGIES assumes no liability for any damages, including
those to third parties, caused by improper use or handling of the instrument.
The user is responsible for connecting the instrument in accordance with the valid
regulations for electrical instruments.
Set the instrument up to ensure easy access to the mains switch.
The mains supply voltage fluctuations must not exceed
+/- 10 % of the nominal voltage. Maximum voltage to be applied is 264 VAC.
The instrument is provided with a 3-pole grounded plug. If your wall outlet does not
allow connection of a 3-pole plug, have a suitable wall outlet installed by qualified
personnel or use an adapter for safe grounding. Please observe the safety specifica-
tions of the grounded plug.
The instruments are designed according to the IEC 61010-1 or EN 610 10-1 regula-
tions for electrical measuring systems.
To disconnect the instrument from power the appliance coupler has to be removed
from the mains. The mains inlet of the instrument is located at the rear panel (see
7.2).
Service and repair work may be carried out by qualified personnel only.
The operator may only perform the maintenance work described in this user guide.
There are no exchangeable electrical components in the instrument. In case of mal-
function call authorized service personnel.
Use only parts described in this manual for servicing.
Pull the power cord to disconnect instrument from power supply.
Turn instrument off before pulling the power cord.
If you can see that the instrument has become unsafe to use, switch it off and dis-
connect it from power supply.
If liquid gets inside the instruments, pull the power cord and remove the flow cell
Clean the measuring chamber or have it cleaned by an authorized service center.
Do not use any flammable or explosive solutions or liquids whose mixture is flamma-
ble or explosive.
The instrument is not intended to be used with biohazardous samples.
The operator is responsible for the use of reagents.
For instrument cleaning, please refer to the respective sections in this manual.
Reliable instrument function can be guaranteed only when original spare parts are
used.
Pollution degree is 2.