9. Parameter Entries
FlowStar2 LB 514
86
Measurement parameters page 1
Run name
Enter the name of the run. This name is displayed on the
Measurement
menu while a measurement is running and it also appears on the printout.
Run time
Enter the duration of the run (max. 24:59:59).
Start mod
e
You can choose
manual
or
external
as start mode. Manual means that
you have to push the function key
Start
on the
Measurement
menu. If
you select
external
, the control takes place via an external device (con-
nect to
“Start” at the
External Control
connector block on the device rear
panel)
Current cell
The parameters for the cell used are displayed here. This entry cannot
be edited. If necessary, you have to replace the measuring cell.
Cell volume [µl]
Cell volume in µl.
Coincidence time [ns]
Is defined by the measuring cell used (decay time of the scintillators):
Solid cells: 100 nsec; Liquid cells: 30 nsec.
Current method
Select the desired method.
Method XXX
Selection and display of the desired method for this run. Push the
button to view the selection list with the methods that are compatible with
the measuring cell used. If no method list is displayed, no method has yet
been defined for this measuring cell.
Method parameters...
Push the
button to display the parameters of the selected method. At
this point, the method parameters cannot be edited.
Liquid handling
These parameters will be displayed only if the LS pump has been ena-
bled on the menu
Configuration | Device
. In this case the measurement
parameters cover 2 pages.
Fraction collector
Tick this checkbox if a fraction collector is employed.