FlowStar2 LB 514
1. Introduction
9
This instrument must not be used in areas with potentially explosive atmosphere.
The instrument may only be used in rooms with a max. pollution degree of 2.
The instrument and the flow cells are designed for flow rates of max. 10ml/min. If
you intend to use higher flow rates, please contact the manufacturer.
The total system pressure must not exceed 200kPa l.
The 12V terminals at the rear connectors do not comply with the requirements of
limited-energy circuits.
The instrument is part of a measuring system with external HPLC-Controller, PC and
/ or other equipment. Hazards resulting from application caused by supplied measur-
ing substances and evaluation of the measured values by external analyses must be
considered in end system / application.
The tests and service work recommended by the manufacturer has to be performed to
make sure that the operator remains safe and that the instrument continues to work
correctly. Any service and maintenance work not described in this user guide has to be
performed by authorized service personnel. Use the instruments only for the designated
application.
Service address:
Berthold Technologies GmbH & Co. KG
Calmbacher Str. 22
D-75323 Bad Wildbad - Germany
Tel. +49 (7081) 177-111
Email: service@berthold.com