17
4.
Software Functions and System Configuration
Depending on the selected configuration, some input dialog boxes are dimmed. If e.g.
“Density measurement without temperature compensation” has been selected, the input
dialog box for temperature compensation is not available.
Chapter 3.4.5 includes an overview of the menu structure.
The factory setting is listed in the
Configuration Checklist
in the Appendix. Enter the final
operating settings in this list.
Individual options are selected by pushing the
<^^^>
button. Push
<enter>
to accept
the selected value.
After the last entry, the instrument has to be connected to mains for at least
5 minutes to ensure that all values entered have been stored in the FLASH
memory and will not be lost if power failure occurs.
4.1
General Data
Password
:
You can enter a 6-digit number. Upon confirmation with <
enter
>, the system is pro-
tected against unauthorized manipulation of the parameters.
Display: Keyboard locked.
All parameters can still be viewed, however. Password protection is revoked as soon as
you enter the correct numerical value and confirm your entry with <
enter
>.
Date
:
Enter the current date in the format TT.MM.YY. The correct date is important for auto-
matic correction of the source activity decay.
Time
:
Enter the current time in the format HH.MM. Time deviations have hardly any effect on
the activity decay correction.
System/Version
:
Display of instrument type and software program version. Please have this data handy if
you need to consult with the manufacturer.
Detector Software:
Display of the software version installed in the detector. Please have this data handy if
you need to consult with the manufacturer.
Language
:
German, English or French can be chosen as dialog language by pressing the <
^^^
> key.
Print Parameter
?
A printer with serial interface can be connected on the front panel. Push
<enter>
to
print all parameters for documentation. You may also connect a PC and transfer the date
to a PC for processing using a terminal program. See also 4.2.
Factory setting
:
Push <
sk1
> to reset all parameters to the factory setting.
The current calibration data is lost when the instrument is reset!
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...