04/03
347
LB 444
23
4.5
Calibrate
Product Select
No. 1/2/3/4:
Select product to be calibrated and confirm with <
enter
>.
Suspension measure
Yes/No
If you select yes the Solid density and liquid density parameters are taken into account in
the calculation (see also chapter 8.4).
Solid density
Enter Solid density (g/cm
3
) of component to be measured (only for suspension meas-
urements). See also chapter 8.4.
Liquid density
Enter liquid density (g/cm
3
) only for suspension measurements. This is the density of the
carrier liquid or, with liquid mixtures, the density of the attendant component.
Water TC ?
Yes/No
Only for measurements of suspensions using water as carrier liquid and temperature
compensation (via PT 100 or current input).
When temperature compensation for water is selected, the automatic temperature com-
pensation (TC1, TC2) is disabled and the temperature coefficient for water is used. Thus,
•
the different temperature coefficient of water at different temperatures and
•
the temperature influence which decreases with rising solids contents
are taken into account.
In this case, the next query is “Reference temperature” (see also chapter 7).
Temperature coefficient TC1
Enter linear temperature coefficient (999.999e
-99
to 999.999e
+99
). This coefficient usually
suffices for minor temperature changes of ± 20°C (see chapter 7).
Temperature coefficient TC2
Enter square temperature coefficient (999.999e
-99
to 999.999e
+99
). We recommend en-
tering TC2 in addition for minor temperature variations > 20°C (see chapter 7).
Reference temperature
Enter the reference temperature for the density, concentration or mass flow measure-
ment. If the actual temperature deviates from the reference temperature entered here,
the measured density or concentration values are temperature corrected (see chapter 7).
Unit select
g/cm
3
, t/m
3
, Bx, Weight%, °Be, g/l, mm, g/m
2
Select the unit for density and concentration measurements and area weight measure-
ments (
mm, g/m
2
)
.
Data input
Enter or read in the calibration data for the requested number of calibration points.
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...