04/03
347
LB 444
29
The distance from the bend must be the larger
•
the higher the flow rate
•
the bigger the difference between liquid density and Solid density.
Measuring system installed outdoors have to be protected from rain and direct sunshine
(e.g. by a thin sheet metal). Please keep in mind that the cables must not be exposed to
temperatures exceeding 70°C. Moreover, cable bushing and cable have to be aligned
such that no water can flow along the cable into the bushing.
RADIOACTIVE
Figure 12: Outdoor installation
On S or U-shaped measuring paths the shielding
container with source has to be installed on top
and the detector at the bottom.
A temperature insulation consisting of glass or
rock wool is not suitable, as it does not ensure
safe installation of the measuring system. If, for
technical reasons, temperature insulation at the
measuring point is absolutely required, it must be
made of a hard, non-hygroscopic material, i.e.
aluminum silicate or it must not be installed di-
rectly on the pipeline, but on separate supports
(see Figure 14).
RA
DI
OA
CT
IV
E
Figure 13: Installation on S or
U-shaped measuring path
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...