04/03
347
LB 444
31
a) Side view
b) Top view
Figure 16: Installation in a container with horizontal flow
a) Side view
b) Top view
Figure 17: Installation in a container with vertical flow
For installation in a container, keep in mind:
Selection of measuring point. At the measuring point, the pipeline must always be com-
pletely filled with the product to be measured. If the pipeline is only partially filled, this
may result in incorrect measurements.
For outdoor installation, the measuring system must be protected from rain and direct
sunshine (e.g. by a thin sheet metal).
Do not expose the cables to temperatures exceeding 70°C.
Protect the detector against temperatures > 50°C by a water cooling jacket.
Cable bushing and cable have to be aligned so that no water can flow along the cable into
the bushing.
The detector installation site must not be subject to heavy vibrations as this could dam-
age the detector.
The product temperature for temperature compensation must be measured directly next
to the density measurement.
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...