04/03
347
LB
444
40
Shielding Container for Installation in a Container
Version with knurled nut
•
Open padlock and turn pro-
tective cap until it can be ta-
ken off.
•
Open knurled nut and pull
source forward to “OPEN” po-
sition.
•
Turn counter nut completely
back toward screw head.
Firmly tighten knurled nut.
•
Firmly turn counter nut to-
ward the front, in the direc-
tion of the protection pipe
and tighten it firmly.
•
Attach protection cap again
and secure it with a padlock.
Figure 23:
Locking mechanism with knurled nut
Version with spring pin
•
Open padlock and turn pro-
tective cap until it can be ta-
ken off. Open fastening
screw.
•
Pull out spring pin and push
source locking lever toward
the front in “OPEN” position
until the spring pin engages.
•
Firmly tighten locking screw
again.
•
Attach protection cap again
and secure it with a padlock.
Shielding with pneumatic locking
drive:
•
Open pressure supply.
CLOSED
OPEN
Spring pin
Locking screw
Figure 24: Locking mechanism with spring pin
Turn instrument power supply on
. Manufacturer’s name, version number and device
type are displayed.
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...