41
6.2.1
Basic Settings
These steps describe how to take the system into operation and how to change the basic
system settings.
Note:
To enter numbers, select the entry position with <
←←←
> and the number you
want with <
^^^
>. Confirm your selection or new entry with <
enter
>.
Press <
run
> to return to the display mode directly any time (except in Service menu).
If no entry is made, the system automatically switches from the current menu item to
the display mode after several minutes.
Important:
After data input the system has to remain turned on for at least 5 minutes to ensure that
all entered data has been saved.
1. Push
<
more
> to select menu
General Data
/
Operating Mode
.
Push <
sk1
> to select
General Data
and <
more
> to call parameters sequentially.
Enter
password
and unlock or lock system with <
enter
>
Check
date & time
and correct it, if necessary
Check
instrument version
(e.g. LB 444, version 2.0)
Select
language
Print Parameter,
only if needed
Factory Setting,
only if needed (see also Software Functions and System Configuration)
Push <
done
> to return to menu group.
2. Push
<
sk2
> to select
Operating Mode
menu and
<more>
to select parameters.
Select
Config Instrument
with
<
^^^
> and confirm with <
enter
>:
Density measurement with or without TC (Pt 100 or current input)
Mass flow measurement with or without TC (Pt 100)
Select
radiation measure
to irradiate the product. Select
backscatter measure
if source
and detector are installed parallel.
Define
Error mode
Select parameters for
RS232 interface
Push <
done
> to return to menu group.
3. Push
<
more
> to select menu
Parameter / Product Data
.
3.1 Push <
sk1
> to select
Parameter
menu and <
more
> to call parameters sequentially.
Select product
Select
detector
and
isotope
Measuring path
in product (e.g. pipe diameter) in cm
Select
Measure mode
(i.e. continuous)
Enter
time constant
(e.g. 60 s)
Disable or enable
rapid switchover
Disable or enable
Interference Radiation
Define
max. count rate
(e.g. 100000 cps)
Define
min. count rate
(e.g. 0 cps)
For TC via current input:
Select
current input
range 0-20mA or 4-20mA
Enter
temperature value
for 0/4mA and 20mA
Push <
done
> to return to menu group.
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...