6.4
Error Messages
6.4.1
Error Messages Reset
All error messages must be reset with <
enter
>.
Several simultaneously or consecutively occurring errors are stored in an error register in
the order of their appearance; they have to be reset individually by pushing <
enter
>
several times.
All errors are signaled via the “Error” output (Relay 16a/16c).
6.4.2
Error Messages during Operation
Error no. Message
Cause, Notes
Remedy
32 Wrong
password
The wrong password has been
entered to unlock the keyboard.
Push “enter” to reset error
message. Enter correct
password.
6.4.3
Error Messages during Calibration
Error no. Message
Cause, Notes
Remedy
10 Not
calibrated
Calibration incomplete
Perform calibration
36 Missing
ca-
lib. points
Not enough points entered for
selected calibration mode
Enter further calibration
points or select another
calibration mode
37 Curve
not
clear
System could not calculate a
clearly rising or falling curve
Check calibration points for
input errors.
Select calibration mode
“linear”.
38 Error
one-
point cali-
bration
Absorption coefficient = 0
Absorption coefficient = 0
39 Measuring
path not
defined
Measuring path = 0
Enter measuring path
56 Range
overflow
Measuring path = 0
Enter measuring path
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...