6.4.4
Error Messages during Measurement
Error
no.
Message Cause,
Notes
Remedy
1
Rate Overflow
Count rate > 520000 cps,
with large pipelines,
pipeline possibly empty
Replace detector, stop
measurement while
emptying pipeline, e.g. via
digital input
2
No Detector Rate
Detector faulty
Replace detector
3
Wrong HV
Radiation exit channel of
shielding container closed,
detector faulty.
Open radiation exit
channel, replace detector
4
Pt 100 wrong
temperature
Cable detector – Pt 100
interrupted, detector faulty,
Pt 100 faulty
Check cable, replace
detector or Pt 100
5
4 –20 mA faulty
Input current below 4 mA,
evaluation unit faulty.
Check input current,
replace evaluation unit.
6 Detector
temperature > 65°C
Detector temperature too
high, detector faulty
Cool detector, if necessary,
replace it
7 Trouble
with
detector
communication
Cable detector – evaluation
unit interrupted or too long
or cable resistance too
high, detector faulty,
evaluation unit faulty
Replace cable or use bigger
wire cross-section
Replace evaluation unit.
8 Measurement
stopped
Measurement process
stopped via digital input
or
measurement stopped
because count rate
threshold has been
exceeded or not reached.
Open digital input.
Disable count rate
thresholds to identify fault.
9 Detector
temperature > 80°C
Detector temperature too
high, detector faulty
Replace detector
10 Missing
calibration Measurement was started
without calibration
Perform calibration
11 Power
fail
> 1 month
Check date after long
period of power failure
If this error message
appears already after a
brief period of power
failure: replace Li–battery.
12
No input allowed
Input locked via password Enter password
47
Rate < Minimum
Enter actual count rate
< than entered min. count
rate:
shielding container closed?
detector fault?
Replace detector
Open shielding container
48
Rate > Maximum
Enter actual count rate
> than entered max. count
rate:
detector fault?
pipeline empty and
shielding container not
closed?
Replace detector
Close shielding container
Summary of Contents for LB 444
Page 1: ...Density Meter LB 444 Id No 32816BA2 Rev No 04 22 04 03 ...
Page 2: ......
Page 9: ...04 03 347 LB 444 VII ...
Page 10: ......
Page 109: ...04 03 347 LB 444 99 13 4 Mounting Device 90 for Pipe Diameter 21 3 76 1 mm ...
Page 112: ...04 03 347 LB 444 102 ...
Page 113: ...04 03 347 LB 444 103 ...
Page 114: ...04 03 347 LB 444 104 ...
Page 117: ...04 03 347 LB 444 1 14 Ex Certificates for Evaluation Unit ...
Page 118: ...04 03 347 LB 444 2 ...
Page 119: ...04 03 347 LB 444 3 ...
Page 120: ...04 03 347 LB 444 4 15 Ex Certificate for the Detectors ...
Page 121: ...04 03 347 LB 444 5 ...
Page 122: ...04 03 347 LB 444 6 ...
Page 123: ...04 03 347 LB 444 7 ...
Page 124: ...04 03 347 LB 444 8 ...